System Manual

26-0002-0003/5 (en) (2021) 60
Outer Material
PMMA plastic (plexiglas - black)
Size
~10 cm × 7.5 cm × 2.5 cm; no sharp corners
Cable Length
~2 m
Weight
~131 g
Connector Type
USB
Communication Distance
≤1 m
Communication Band
Medical Implantable Band (MICS, ~403 MHz)
Table 16.1: Characteristics of the Wireless Transmitter
16.4 Stroke Application & Programming Software (SAPS)
The SAPS software was developed by MicroTransponder to manage IPG therapy and therapy
information. The SAPS has the following capabilities:
Allows the physician to set the IPG stimulation parameters
Allows the physician to check the status of the IPG battery level
Allows the physician to check the Lead impedance
Allows the physician to record the rehabilitation task information
Stores the therapy session history for review at a later time
Requires the physician to use a SAPS password to access implant programming and
therapy functions
Allows users without the SAPS password (but with the laptop password) to access patient
log and therapy information
The software is delivered to the physician pre-installed on a commercially available laptop.
See Section 3: Using the Software (SAPS) for more information on use of the SAPS.
16.5 Other Applicable MicroTransponder
®
Products
Model 1001 IPG
Model 3000 Lead
16.6 Applicable Commercial Devices for Use with MicroTransponder IPG
A commercial magnet with a pull force rated at 4.5 kg or greater is sufficient to close the reed
switch within the IPG and stop stimulation.