The Wireless Transmitter is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Wireless Transmitter can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Wireless Transmitter as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
M
150 kHz to 80 MHz
𝑑 = 1.2
√
𝑃
80 MHz to 800 MHz
𝑑 = 1.2
√
𝑃
800 MHz to 2.5 GHz
𝑑 = 2.3
√
𝑃
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.2
1.2
2.3
10
3.7
3.7
7.4
100
12
12
23
Table 17.4: Recommended Separation Distance Between Portable and Mobile RF
Communication Equipment and the Wireless Transmitter
Telemetry Description
The Model 1001 IPG and Model 2000 Wireless Transmitter communicate with each other using
ultra-low power Medical Implant Communications Service (MICS Band) telemetry in the
frequency range 402 – 405 MHz.