ABORT LEVEL LOW
If the liquid level sensor is set to ON and the liquid drops below the required level during
operation, the cleaning cycle will stop and the following screens will alternate.
Solutions:
Fill the QuickClean™ with water so that it is above the fill line indicated inside the tank.
If you are using “normal” water ensure that the detergent has been added.
If the water in your area is “hard”, slightly more detergent may be required.
If “pure” water (e.g. De-I or RO) is being used, the level sensor can be disabled in the advanced
menu (see page 20).
USER CYCLE ABORT
If the operator aborts the cycle, the following screen will be displayed
Troubleshooting and Aborting a Cycle
If the unit fails to turn on (no display is shown) Check that the unit is plugged in and that
main power is present.
If a cycle is aborted, by the operator or by the unit, the reason for the failure will be dis-
played.