Instruction manual

DC Power Requirement
External DC voltage power supply. This power supply must be ground isolated in such a
way that the DC voltage output floats with respect to ground.
Maximum voltage limit 300 VDC
Maximum power limit 50 W
Size and Weight
Overall size Lid and Tank Assembly: 39 cm (L) x 20 cm (W) x 42 cm (H)
Shipping weight 16 Kg
Environmental Requirements
Storage environment 0–70 °C, humidity 0–95% (non-condensing)
Operating environment 0–35 °C, humidity 0–95%.
Regulatory
Meets the requirements of IEC 1010-1 and FCC, Class A.
1.3 Description of Major Components
Fig.1.1. 1. Lower buffer tank. 2. Lid with temperature controller. 3. Core. 4. Comb gasket holder.
5. Sandwich clamps. 6. Casting stand with sponge. 7. Stopcock. 8. Combs and spacer set.
9. Alignment card. 10. Model 475 Gradient Delivery System. 11. Filler spacer.
1. Lower Buffer Tank
The lower buffer tank encloses the unit and provides stability during electrophoresis.
Always keep the buffer level between ‘Fill’ and ‘Maximum’ during a run.
2. Lid with Temperature Controller
Combined with the lower buffer tank, the lid acts to fully enclose the system. The lid
should be placed so that the tip of the stirring bar fits inside the support hole at the tank. The
loading lid should be left on the system at all times unless you are loading samples.
3. Core
The completed gel sandwich attaches to the core so that the outer plate of sandwich forms
the side of the upper buffer chamber. The inner plate is clamped against a rubber gasket on
3