Operating instructions
Table Of Contents
- Contents
- Description of the machine 4
- Guide to the machine 6
- Warning and safety instructions 11
- Opening and closing the door 16
- Automatic mobile unit recognition (Optional) 17
- Areas of application 21
- Dispensing liquid chemical agents 30
- Operation 34
- Data transfer 39
- Maintenance 40
- Periodic checks 40
- Process validation 41
- Routine checks 41
- Cleaning the filters in the wash cabinet 42
- Sprüharme reinigen 44
- Cleaning the control panel 45
- Cleaning the front of the machine 45
- Cleaning the wash cabinet 45
- Cleaning the door seals 45
- Protein test 45
- Mobile units, modules and inserts 46
- Printer (optional) 47
- After sales service 48
- Converting the type of heating 49
- Electrical connection 50
- Plumbing 51
- Technical data 52
- Special features - OXIVARIO 53
- Disposal of your old appliance or machine 55

In operating level C the automatic
mobile unit recognition system (AWK)
allocates programme places from 1 - 15
to mobile units with the correct coding.
The unit coding (on the mobile unit) and
the programme place with the
corresponding programme (in the
electronic control unit) have to match
each other.
To do this:
^
Each mobile unit must be coded
before being used for the first time
(see "Coding the mobile unit"), and
^ The programme for which the mobile
unit is coded must be allocated to the
correct programme place.
See "System function - Selector switch
organisation" in the Programming
manual for information on how to
change programme positions.
,
Before starting a programme it is
absolutely essential that you check
that the programme shown in the
display is the correct one for the
mobile unit being used.
Otherwise inadequate cleaning or
disinfection could be the result.
Please make sure, that the places
assigned for the programmes using
Automatic mobile unit recognition
are not changed around arbitrarily.
Automatic mobile unit recognition (Optional)
17