than 2½ cm (1”) below the fill hole, add MA9011 3.5M KCl Electrolyte Solution. For
faster response, unscrew the fill hole screw during measurements.
STORAGE PROCEDURE
To minimize clogging and assure a quick response time, the glass bulb and the junction of
the electrode should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of MA9015 Storage Solution
or, in its absence, Filling Solution (MA9011). Follow the Preparation Procedure before
taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to the instrument must
be intact and there must be no points of broken insulation on the cable or cracks on the
electrode stem or bulb. Connectors must be perfectly clean and dry. If any scratches or
cracks are present, replace the electrode. Rinse off any salt deposits with water.
Probe Maintenance
Refill the reference chamber with fresh electrolyte (MA9011). Allow the electrode to
stand upright for 1 hour. Follow the Storage Procedure above.
CLEANING PROCEDURE
Wine deposits Soak in Milwaukee MA9016 cleaning solution for 1 hour
Wine stains Soak in Milwaukee MA9016 cleaning solution for 1 hour
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thor-
oughly with distilled water, refill the reference chamber with fresh electrolyte and soak the
electrode in MA9015 Storage Solution for at least 1 hour before taking measurements.