Specifications

8
Calibration Procedures: 2238-20
ModeM2238 True RMS Clamp Meter for HVAC/R
EnvironmentalTemperature 18°C ~ 28°C Humidity<70%RH
CC aalliibbrraattiioonn eeqquuiippmmeenntt
11..
FLUKE 5520A 22..FLUKE 5500A/COIL(50TURNS) 33..MM22223388 Fixture
44..TT eemmppeerraattuurree CC oonnnneeccttoorr 55..Screw adjustment probe
NOTE:
(1) Turn the unit off. Press and hold “ + ” keys and turn the function dial
to range “
” then push “°F/°C ” button within 2 sec. The LCD will blink “ZERO”
Step Function Switch Position Input Signal Adjustment Adjust Range Note
1.
DCV Adjustment
300VDC
VR1
299.9~ 300.1V
2.
ACV Zero Adjustment
Open Circuit
VR2
0.0~0.1V
(1)
3.
ACV Adjustment
300V/60Hz (sine wave)
VR3
299.9~300.1V
4.
DCuA Adjustment
300uADC
VR4
299.9~300.1uA
5.
ACA 600A Adjustment
300A/60Hz VR5 299.9~300.1A
6.
ACA 60A Adjustment
30A/60Hz
VR6
29.99~30.01A
7.
Capacitor Adjustment
220uF
Soft Key
218.0~222.0
(2)
8. Capacitor Adjustment
2200uF
Soft Key
2180~2220
(3)
9.
Temp. Adjustment
0mVDC
Soft Key
---
(4)
10. Temp. Adjustment
16.395mVDC
Soft Key
---
(5)
11.
Temp. Adjustment
0°C
Soft Key
-0.1~0.1°C
(6)