Product Info
5. CLEANING, MAINTENANCE AND STORAGE
Please, refer to section 1 Safety Information.
Both, the Physilog6 sensor device and the clip accessories can be manually cleaned and disinfected
carefully using wipes with different chemicals (available at drugstore or pharmacy) detailed under
section 5.1 of the present document.
5.1. Cleaning
Clean and disinfect the Physilog6 sensor device and the clip accessories before the first use at home as
follows:
● Unpair the Physilog6 sensor device from the Mobile Application by closing the Gait Up GO
mobile application prior to cleaning
● Unclick the Physilog6 sensor device from the clip accessories prior to cleaning
● Unplug the Physilog6 sensor device from the USB cable/power prior to cleaning
● Use Hydrogen peroxide 3%, bleach solution 50% or saline solution to clean and disinfect the
Physilog6 sensor device’s clip accessories. Do not use a cleaning spray, but carefully wipe it
using a humid wipe.
● Use 96% ethanol solution or 70% isopropyl solution to clean and disinfect the physilog6 sensor
device’s plastic enclosure. Do not use a cleaning spray, but carefully wipe it using a humid
wipe.
● Liquids and foreign material (e.g. dust) must not be allowed to enter the Physilog6 accessory
device. Do not clean the device with excessive liquid.
● Ensure that the Physilog6 sensor device and the clip accessories are dry prior to being used
again.
5.2. Maintenance
The Gait Up GO medical device is tested and calibrated prior to distribution. Under normal conditions,
the characteristics of the device do not vary.
● The Physilog®6 should be handled carefully.
● If the Gait Up GO medical device (incl. all its accessories) shows functional instabilities or any
visual defects, stop using it and contact the customer service.
The Physilog®6 firmware is updated using the Gait Up GO mobile application provided by Gait Up SA
according to the instructions of use section 4.5.5.
The Physilog®6 sensor device does not include any consumables (one-time use).
The fixation clips used to fix the Physilog6 sensor device on the Participant’s body are spare parts. If
they get damaged, please contact Gait Up SA support to purchase new ones.