Product Info

IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
N/A (DC supply)
Frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m, 50 or 60 Hz
tested in X, Y and Z
axes.
30 A/m, 50 or 60 Hz
Power frequency
magnetic fields should
be at levels
characteristic of a
typical commercial or
hospital environment.
Radiated RF immunity
IEC 61000-4-3
10 V/m @
80MHz-2.7GHz
80% AM à 1kHz
10 V/m @
80MHz-2.7GHz
80% AM à 1kHz
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Physilog6
sensor device,
including the USB
cable, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter
Minimum separation
distance shall be
calculated by following
equation:
E= (6/D)*sqrt(P)
Where:
E is the
immunity test
level in V/m
P is the
maximum
output power
rating of the
transmitter in
watts (W)
according to
the transmitter
ENGLISH
76/84