Table 12. Electromagnetic immunity
Note (a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To access the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the device is used exceeds the applicable RF compliance level above,
the Gait Up GO medical device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocation of
the Gait Up GO medical device.
Note (b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m
9.6 Recommended separation distances between portable and mobile RF
communications equipment bas the Gait Up GO medical device
Conducted RF
immunity
IEC 61000-4-6
3 V 0,15 MHz – 80 MHz
6 V in ISM bands,
between 0,15 MHz and
80 MHz
80 % AM @ 1 kHz
3 V 0,15 MHz – 80 MHz
6 V in ISM bands,
between 0,15 MHz and
80 MHz
80 % AM @ 1 kHz
If the measured field
strength in the
location in which the
Physilog 6
is used exceeds the
applicable RF
compliance level, the
Physilog 6 should
be observed to verify
normal operation. If
abnormal
performance is
observed, additional
measures may be
necessary, such as
reorienting or
relocating the Physilog
6
.
5240 – 5500 -
5785
Modulation
pulsée 217 Hz
9
9