Product Info

9.7 FCC - Information to user
This document, which describes the instructions for use for an intentional radiator, cautions the user
that changes or modifications not expressly approved by the Manufacturer could void the user's
authority to operate the equipment.
[54 FR 17714, Apr. 25, 1989, as amended at 68 FR 68545, Dec. 9, 2003]
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) this device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
Rated maximum
output power of
transmitter
80 KHz to 1000 Mhz
outside ISM bands
1GHz to 2.7GHz
outside ISM bands
2.7GHz to 6GHz
outside ISM bands
d=0.6*sqrt(P)**
d=0.6*sqrt(P)**
d=2*sqrt(P)**
0.01W
0.06
0.06
0.2
0.1W
0.19
0.19
0.63
1W
0.6
0.6
2
10W
1.90
1.90
6.32
100W
6
6
20
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres [m] can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts [W] according
to the transmitter manufacturer.
*The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765-6.795
MHz, 13.553-13.567 MHz, 26.957-27.283 MHz and 40.66-40.7 MHz.
**Formulas coming from the IEC 60601-1-2