4.5.7. Operating a test
Once the new session is started, the selected tests can be performed and the Gait Up GO mobile
application can be operated either by a medical professional, by the Participant in presence of the
medical professional or by the Participant or the patient relative at home without the presence of the
medical professional. The result interpretation must however always be carried out by a medical
professional.
Preparing the Participant
Before attaching the sensor device accessory to the Participant, please check the integrity of the skin
in the area where the sensor will be placed. In case of redness or skin damage, do not fix the sensor on
the Participant.
Make sure that the participant is wearing adequate shoes where the clip can be attached firmly and
without heels.
Check Physilog6 sensor connection and battery level
Check that the Physilog6 is connected to the Gait Up GO mobile application and that it has a sufficient
battery level. Check the sensor status by accessing the sensor information from the button in the top
right corner as shown in Figure 40 below.
If the battery is lower than 25%, the Gait Up GO Mobile Application will not allow it to start any test.
The battery level indicating the battery status of the connected Physilog6 sensor device on the Mobile
Application Screen is shown in Figure 41.
NOTE: the Bluetooth of the mobile phone / tablet needs to be activated. For Android the localisation
must in addition be activated to be able to use Bluetooth low energy. The Gait Up GO application does
not access your localisation information.
NOTE: To “wake up” the Physilog6 sensor from stand-by mode when it is not connected, lightly shake
the device to turn on its BLE (“shake up to wake up”).
NOTE: in case of any issue with Physilog6 connection, check the troubleshooting section 6. If problems
persist, contact the Gait Up SA support team at support@gaitup.com.
Fig 55. Physilog linked Fig 56. Physilog information
Select the test
ENGLISH