User Manual

Table Of Contents
Demographics Tab Patient Setup Functions
5 - 14 0070-10-0690-01 Panorama™ Operating Instructions
2. Rotate the dial control to change the V-Tach Rate. The range available is 100-180 bpm,
in 5 bpm increments.
3. Select the Asystole Delay button to change the asystole delay time for arrhythmia
processing. When using a 3 or 5-wire lead set, the range available is 3-10 seconds.
When using a View 12
card, the range available is 3-8 seconds.
4. Select the V-Tach Threshold button to change the V-Tach threshold settings. The
range available is 3-15 beats.
5. Select the ECG Noise Delay button to change the noise delay setting. The range
available is 3-30 seconds.
NOTE: The ECG Noise Delay button is only displayed and enabled
when using a Telepack device. This button is not available
when using a bedside monitor (shown in FIGURE 5-10). ECG
Noise Delay must be enabled or disabled at the beside
monitor.
6. Accept or reject the selections made in the dialog box.
Select the Done button to accept the selections.
Select the Cancel button to discard the selections, close the dialog box and return to
the Unit Choices tab.
For additional information, refer to ‘‘Simultaneous Patient Demographic Tab Editing’’ on
page 5-17.
5.1.3.3 Configuring Care Groups
Select the Care Group sidebar button to assign a specific care group to a patient tile. Care
groups allow user-defined categorization of patient tiles. Care groups may represent a
care-giver, hospital department, or a special condition. Each care group button has a color
indicator that, when assigned to a patient, appears to the right of the VIEW button in the
patient tile. The system default settings for Care Groups are defined in the System Care
Group tab. For additional information, refer to the “Care Group Tab” on page 9-9.
The Care Group tab and the associated colorized indicators are not available at the
ViewStation or the WorkStation.
To select a care group:
1. From the Demographics tab, select the Care Group sidebar button. The Care
Group dialog box (shown in FIGURE 5-11) is displayed.
FIGURE 5-11 Care Group Dialog Box
Care Group Button