Panorama™ Operating Instructions 0070-10-0690-01 5 - 15
Patient Setup Functions Demographics Tab
2. Select a Care Group button.
3. Accept or reject the Care Group selection.
•The Done button accepts the care group selection.
•The Cancel button discards the selection, closes the dialog box, and returns to the
Demographics tab.
5.1.3.4 Settings (Telepack)
Select the Settings sidebar button to establish the type of ECG monitoring filter that will be
used for the Telepack device at the central station or the WorkStation. There are two types of
monitoring available:
NOTE: The Settings sidebar button is not available at the
ViewStation.
•The Monitor filter establishes the frequency response range for non-ST ECG monitoring.
•The ST filter establishes the frequency response range for ST diagnostic monitoring.
To enter settings for a Telepack:
1. From the Demographics tab, select the Settings sidebar button. The Settings dialog
box (shown in FIGURE 5-12) is displayed.
FIGURE 5-12 Settings Dialog Box
2. Select the ST (optional) check box to enable or disable ST analysis for the selected
patient.
A checkmark in the ST box indicates that ST analysis is enabled.
An empty box indicates that ST analysis is disabled.
When the ST check box is empty, the Monitor filter is automatically enabled.
NOTE: If ST licenses are available, an ST license is automatically
assigned to the patient when ST is enabled
.
NOTE: When using a 3-lead set, ST is performed on the lead being
viewed.
3. Select the Monitor or ST filter mode.
NOTE: One filter box must always be populated. When a filter
mode is selected, the other mode is automatically disabled.
A checkmark in the Monitor filter box indicates that the Monitor filter is enabled and
ST analysis is disabled.