User Manual
Table Of Contents
- 08-0308 - Manual Cover.pdf
- 0070-10-0690-01 Panorama Operations Maual with Telepack 2.4 - CH16 REVISED2.pdf
- Foreword
- Warnings, Precautions and Notes
- Indications for Use
- Unpacking Information
- Symbols and Descriptions
- Technical Support
- 1.0 System Overview
- 1.1 Components of the Panorama System
- 1.2 Licensing Key
- 1.3 Supported Devices
- 1.3.1 Panorama Central Station
- 1.3.2 Panorama ViewStation and WorkStation
- 1.3.3 Panorama Telemetry Server
- 1.3.4 Panorama Wireless Transceiver
- 1.3.5 Panorama Gateway
- 1.3.6 Panorama Web Viewer
- 1.3.7 Passport 2® Bedside Monitor
- 1.3.8 Spectrum™ Bedside Monitor
- 1.3.9 Spectrum OR™ Bedside Monitor
- 1.3.10 Panorama Telepack
- 1.3.11 Network Printer
- 1.4 Troubleshooting
- 2.0 Getting Started
- 3.0 Alarm Behavior
- 4.0 Arrhythmia and ST Analysis
- 5.0 Patient Setup Functions
- 5.1 Demographics Tab
- 5.2 Transfer Tab
- 5.3 Discharge Tab
- 5.4 Display Tab
- 5.5 Standby Tab
- 5.6 Wave Gain Tab (Patient)
- 5.7 Print Setup Tab (Patient)
- 6.0 Patient Alarm Functions
- 7.0 Views
- 7.1 Bedside Tab
- 7.2 Trends Tab
- 7.3 Events Tab
- 7.4 Disclosure Tab
- 7.4.1 Accessing the Disclosure Tab
- 7.4.2 Disclosure Tab (Waveforms View)
- 7.4.2.1 Accessing the Waveforms View
- 7.4.2.2 Components of the Waveforms View
- 7.4.2.3 Navigating in the Waveforms View of the Disclosure Tab
- 7.4.2.4 ST Button
- 7.4.2.5 Comments Dialog Box (Optional)
- 7.4.2.6 Sidebar Buttons
- 7.4.2.7 Printing a Full Disclosure Zoom In Report from the Waveforms View
- 7.4.2.8 Troubleshooting
- 7.5 ST Tab
- 8.0 Report Functions
- 8.1 Patient Reports Tab
- 8.2 System Reports Tab
- 8.3 Print Status Tab
- 8.4 Additional Reports
- 9.0 System Setup Functions
- 9.1 Parameter Color Tab
- 9.2 Touch Screen Tab
- 9.3 Installation Setup Tab
- 9.4 Care Group Tab
- 9.5 Volume Tab
- 9.6 System Alarms Tab
- 9.7 Passwords Tab
- 9.8 Equipment Setup Tab
- 9.9 More Tab
- 9.10 Previous Tab
- 9.11 Date/Time Tab
- 9.12 Print Setup Tab (System)
- 9.13 Unit Priorities Tab
- 9.14 Unit Choices Tab
- 9.15 Wave Gain Tab (System)
- 9.16 Wireless Tab
- 10.0 Telepack (Optional)
- 11.0 User Maintenance
- 12.0 Parameter Specifications
- 12.1 Parameter Specifications
- 12.1.1 Heart Rate: HR
- 12.1.2 Respiration: Resp
- 12.1.3 NIBP
- 12.1.4 SpO2
- 12.1.5 Nonin SpO2
- 12.1.6 IBP: mmHg
- 12.1.7 CO2
- 12.1.8 O2
- 12.1.9 INSP/ET O2
- 12.1.10 N2O
- 12.1.11 Anesthetic Agents: Iso, Enf, Des, Sev, Hal, Agent
- 12.1.12 Premature Ventricular Contraction (PVC)
- 12.1.13 Temp: T1,T2, Delta T, and T Blood
- 12.1.14 ST
- 12.1.15 Cardiac Output
- 12.1.16 Cardiac Index
- 12.1.17 Pulmonary Artery Wedge Pressure (PAWP)
- 12.1.18 Battery Percentage
- 12.2 Invalid Data Display
- 12.1 Parameter Specifications
- 13.0 Alarms and Events
- 14.0 Network Tab (Optional)
- 14.1 Network Tab
- 14.1.1 Accessing the Network Tab
- 14.1.2 Sorting the Contents of the Network Tab List Box
- 14.1.3 Assigning Patient Tiles in the Network Tab
- 14.1.4 Removing a Patient Tile Assignment in the Network Tab
- 14.1.5 Removing All Patient Tile Assignments in the Network Tab
- 14.1.6 Refreshing the Network Tab List Box
- 14.1.7 Troubleshooting
- 14.1 Network Tab
- 15.0 Paging Tab (Optional)
- 15.1 Paging Password Tab
- 15.2 Paging Tab
- 15.2.1 Accessing the Paging Tab
- 15.2.2 Adding a Pager
- 15.2.3 Assigning a Patient Tile
- 15.2.4 Adding All Patient Tiles to a Pager
- 15.2.5 Removing a Patient Tile Assignment
- 15.2.6 Removing All Patient Tiles from a Pager
- 15.2.7 Editing a Pager
- 15.2.8 Removing a Pager
- 15.2.9 Resending a Failed Page
- 15.2.10 Generating a Paging Report
- 15.2.11 Troubleshooting
- 17.0 Glossary
Demographics Tab Patient Setup Functions
5 - 20 0070-10-0690-01 Panorama™ Operating Instructions
5.1.5 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Demographics tab.
MESSAGE/ISSUE * REASON SOLUTION
Can not admit Patient to
this tile. No Device
attached to this tile
There is no device assigned to
the selected tile.
Select the OK button to close
the dialog box. Select a patient
tile that has a device attached
to it or attach a device to the
selected patient tile.
The Arrhythmia sidebar
button is disabled.
No equipment is assigned to
the patient tile.
Select a different patient tile.
The selected patient is in the
Standby mode.
Remove the selected patient tile
from Standby mode.
Monitoring device is in
Communications Lost state.
Restore communication to
central station.
The Arrhythmia Menu at the
bedside monitor is active.
Wait for the bedside monitor
Arrhythmia Menu to time out or
be closed or, use the bedside
monitor to make the necessary
arrhythmia changes.
The Pacer, Care Group, and
the Retrieve Discharged
Pt. sidebar buttons are
disabled.
The selected patient tile is in
Standby mode.
Remove the selected patient tile
from Standby mode.
The message These settings
are currently being edited
remotely is displayed.
Select the OK button to close
the message box and enable
the sidebar button.
The filter mode cannot be
Monitor when ST is on
Monitor mode cannot be
selected while ST is enabled.
Select the OK button to close
the dialog box. Turn off ST
select the correct filter mode.
Turning on ST will change
your ECG filtering to ST
mode
ST mode was selected while in
Monitor mode.
Select the OK button to close
the dialog box. Select the
correct filter mode.
Retrieve Discharged
Patient Failed: The
destination tile is
currently occupied.
A patient cannot be retrieved
into an occupied tile.
Select the OK button to close
the dialog box. Select a vacant
patient tile before attempting to
retrieve a patient.
No demographic data is
displayed at ViewStation or
WorkStation for the selected
patient tile.
The host central station is in a
Central Communications Lost
mode.
Resolve the communication
error at the host central station.
Patient tile does not contain a
care group indica
tor
If viewing patient tile from the
central station, a care group
has not been indicated.
Assign the patient tile to a care
group.
Patient tile is being viewed
from a ViewStation or a
WorkStation, which does not
support this feature.
To discern if the patient is
assigned to a care group, view
the patient tile from the host
central station.
These settings are
currently being edited
remotely.
The Care Group dialog box
was closed while the bedside
device’s Patient Menu is active.
Select the OK button to close
the message box and enable
the sidebar button.
* Messages are shown in all bold text.