User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 5 - 23
Patient Setup Functions Transfer Tab
When a patient tile actively monitoring a patient, but experiencing a Communications
Lost event, is selected as the source
To perform a Room To Room Transfer Without Device:
1. Select the patient tile to be transferred. Choose a patient tile by selecting its digital data
or waveform area.
2. Select the Room To Room Without Device button. The following prompt is
displayed: Please touch tile to which you wish to transfer the patient.
3. Select a destination tile that has a device assigned. The message All of the patient's
data will be deleted in the destination tile. Do you still want to continue
with the transfer? is displayed.
5.2.2.1 Effects of the Transfer
The following is a list of the changes that occur at the destination tile when a Room To
Room Transfer Without Device is complete:
Display
The transferred patient's parameter display changes only when the destination device
does not support the source device parameters.
The transferred patient's parameter display waveform 1 changes to ECG II when the
destination device does not support the source device's waveform 1 parameter and has a
5 lead ECG cable.
The transferred patient's parameter display waveform 1 changes to the active lead when
the destination device does not support the source device's waveform 1 parameter and
has a 3 lead ECG cable.
•The Yes button initiates the transfer. As the transfer proceeds, there is no
indication in the Transfer tab.
If the destination tile is in Standby mode when the transfer is complete, the
digital data labels and the demographic information will be displayed in the
patient tile, but the digital data values and waveforms will not be displayed until
the device is returned to the normal monitoring mode.
If the destination tile is in a Communications Lost state when the transfer is
complete, the tile is blank and all buttons on the Transfer tab are unavailable.
The tile remains blank until communication has been re-established.
When the transfer is complete, the source tile is blank and the destination tile
remains blank until the patient is connected to the monitoring device.
If the destination tile is in a Waiting for Patient state when the transfer is
complete, the tile will display the Waiting for Patient message until the
destination device is activated.
NOTE: When the transfer is complete, a Room Transfer
event and an Admit event are recorded in the Event
database.
NOTE: After a Room To Room Transfer Without Device, the
patient’s Paging settings will remain the same.
• The No button terminates the transfer and returns to the Transfer tab.