User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 5 - 31
Patient Setup Functions Transfer Tab
5.2.5 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Transfer tab.
Room To Room Transfer
MESSAGE/ISSUE * REASON SOLUTION
Patient tile shown with
demographic text and
digital data labels, but no
waveforms and/or digital
values are displayed
The selected tile is still in the
Bedside and Panorama
Standby mode.
Return the monitoring device to its
active monitoring state.
The Room To Room
Without Device button
is disabled.
The Transfer Failed list is full.
The list stores a maximum of 5
failed transfers.
Delete a patient from this list before
transferring.
A patient tile is not selected
on the Main Screen.
Select the patient tile that is intended
to be transferred.
The selected patient tile has a
device assigned but is not
monitoring a patient.
Select the patient tile that is intended
to be transferred.
The selected patient tile has
no device assigned.
Select the patient tile that is intended
to be transferred.
The Failed Patient List
is full. Delete a
patient from this list
before transferring.
The Transfer Failed list is full.
The list stores a maximum of 5
failed transfers.
Delete a patient from this list before
transferring.
Unable to select a patient
tile on the Panorama
Display.
A message box is currently
being displayed.
Close the message box or wait for it
to time out and close automatically.
The destination and
source tile cannot be
the same. Please try
again.
The original source tile was
selected when the prompt
“Please touch tile to which you
wish to transfer the patient.”
was displayed.
Restart the transfer process and
select an appropriate destination tile
when prompted. Do not select the
original source tile.
The destination tile
must have a device
assigned to perform
the transfer.
The destination tile does not
have a device assigned.
Restart the transfer process and
select a destination tile that has a
device assigned when prompted.
This Tab is in use by
another display. Can
not be updated from
more than one
display at the same
time.
An attempt was made to open
the Transfer tab on one
display when it is already
open on the other display.
Select the OK button and use the
open Transfer tab.
At the destination tile, the
digital data is displayed
but the waveform is not
displayed
The destination tile was in
Standby mode when the
transfer completed.
Return to normal monitoring mode.
Room to Room Transfer
Without device failed
Another tab was selected
during the transfer process.
Restart the transfer process and do
not select another tab until the
transfer process is complete.
* Messages are shown in all bold text.