User Manual
Table Of Contents
- 08-0308 - Manual Cover.pdf
- 0070-10-0690-01 Panorama Operations Maual with Telepack 2.4 - CH16 REVISED2.pdf
- Foreword
- Warnings, Precautions and Notes
- Indications for Use
- Unpacking Information
- Symbols and Descriptions
- Technical Support
- 1.0 System Overview
- 1.1 Components of the Panorama System
- 1.2 Licensing Key
- 1.3 Supported Devices
- 1.3.1 Panorama Central Station
- 1.3.2 Panorama ViewStation and WorkStation
- 1.3.3 Panorama Telemetry Server
- 1.3.4 Panorama Wireless Transceiver
- 1.3.5 Panorama Gateway
- 1.3.6 Panorama Web Viewer
- 1.3.7 Passport 2® Bedside Monitor
- 1.3.8 Spectrum™ Bedside Monitor
- 1.3.9 Spectrum OR™ Bedside Monitor
- 1.3.10 Panorama Telepack
- 1.3.11 Network Printer
- 1.4 Troubleshooting
- 2.0 Getting Started
- 3.0 Alarm Behavior
- 4.0 Arrhythmia and ST Analysis
- 5.0 Patient Setup Functions
- 5.1 Demographics Tab
- 5.2 Transfer Tab
- 5.3 Discharge Tab
- 5.4 Display Tab
- 5.5 Standby Tab
- 5.6 Wave Gain Tab (Patient)
- 5.7 Print Setup Tab (Patient)
- 6.0 Patient Alarm Functions
- 7.0 Views
- 7.1 Bedside Tab
- 7.2 Trends Tab
- 7.3 Events Tab
- 7.4 Disclosure Tab
- 7.4.1 Accessing the Disclosure Tab
- 7.4.2 Disclosure Tab (Waveforms View)
- 7.4.2.1 Accessing the Waveforms View
- 7.4.2.2 Components of the Waveforms View
- 7.4.2.3 Navigating in the Waveforms View of the Disclosure Tab
- 7.4.2.4 ST Button
- 7.4.2.5 Comments Dialog Box (Optional)
- 7.4.2.6 Sidebar Buttons
- 7.4.2.7 Printing a Full Disclosure Zoom In Report from the Waveforms View
- 7.4.2.8 Troubleshooting
- 7.5 ST Tab
- 8.0 Report Functions
- 8.1 Patient Reports Tab
- 8.2 System Reports Tab
- 8.3 Print Status Tab
- 8.4 Additional Reports
- 9.0 System Setup Functions
- 9.1 Parameter Color Tab
- 9.2 Touch Screen Tab
- 9.3 Installation Setup Tab
- 9.4 Care Group Tab
- 9.5 Volume Tab
- 9.6 System Alarms Tab
- 9.7 Passwords Tab
- 9.8 Equipment Setup Tab
- 9.9 More Tab
- 9.10 Previous Tab
- 9.11 Date/Time Tab
- 9.12 Print Setup Tab (System)
- 9.13 Unit Priorities Tab
- 9.14 Unit Choices Tab
- 9.15 Wave Gain Tab (System)
- 9.16 Wireless Tab
- 10.0 Telepack (Optional)
- 11.0 User Maintenance
- 12.0 Parameter Specifications
- 12.1 Parameter Specifications
- 12.1.1 Heart Rate: HR
- 12.1.2 Respiration: Resp
- 12.1.3 NIBP
- 12.1.4 SpO2
- 12.1.5 Nonin SpO2
- 12.1.6 IBP: mmHg
- 12.1.7 CO2
- 12.1.8 O2
- 12.1.9 INSP/ET O2
- 12.1.10 N2O
- 12.1.11 Anesthetic Agents: Iso, Enf, Des, Sev, Hal, Agent
- 12.1.12 Premature Ventricular Contraction (PVC)
- 12.1.13 Temp: T1,T2, Delta T, and T Blood
- 12.1.14 ST
- 12.1.15 Cardiac Output
- 12.1.16 Cardiac Index
- 12.1.17 Pulmonary Artery Wedge Pressure (PAWP)
- 12.1.18 Battery Percentage
- 12.2 Invalid Data Display
- 12.1 Parameter Specifications
- 13.0 Alarms and Events
- 14.0 Network Tab (Optional)
- 14.1 Network Tab
- 14.1.1 Accessing the Network Tab
- 14.1.2 Sorting the Contents of the Network Tab List Box
- 14.1.3 Assigning Patient Tiles in the Network Tab
- 14.1.4 Removing a Patient Tile Assignment in the Network Tab
- 14.1.5 Removing All Patient Tile Assignments in the Network Tab
- 14.1.6 Refreshing the Network Tab List Box
- 14.1.7 Troubleshooting
- 14.1 Network Tab
- 15.0 Paging Tab (Optional)
- 15.1 Paging Password Tab
- 15.2 Paging Tab
- 15.2.1 Accessing the Paging Tab
- 15.2.2 Adding a Pager
- 15.2.3 Assigning a Patient Tile
- 15.2.4 Adding All Patient Tiles to a Pager
- 15.2.5 Removing a Patient Tile Assignment
- 15.2.6 Removing All Patient Tiles from a Pager
- 15.2.7 Editing a Pager
- 15.2.8 Removing a Pager
- 15.2.9 Resending a Failed Page
- 15.2.10 Generating a Paging Report
- 15.2.11 Troubleshooting
- 17.0 Glossary
Alarm Limits Tab (Patient) Patient Alarm Functions
6 - 10 0070-10-0690-01 Panorama™ Operating Instructions
6.2.7 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
patient Alarm Limits tab.
MESSAGE/ISSUE * REASON SOLUTION
No Patient Selected! This message appears in the
demographic field when either
no patient tile or an empty
patient tile has been selected.
Select an active patient tile.
Communications Lost Communication between the
Panorama Central Station and
the monitoring device for the
selected patient tile has been
interrupted.
Verify the connection between
the Panorama Central Station
and the monitoring device. For
example, is the monitoring
device out of range?
All of the sidebar buttons
(except the Normal Screen
button) are disabled.
An empty patient tile is
selected.
Select a valid patient tile by
touching the digital or
waveform area of the patient
tile.
All Alarms Suspended The Suspend All Alarms
function is enabled. All numeric
physiological, lethal
arrhythmia, non-lethal
arrhythmia and technical alarm
processing for the selected tile
has been suspended.
Resume alarm processing by
selecting the Suspend All
Alarms sidebar button and
disabling the Suspend All
Alarms function.
Non-Lethal Alarms
Suspended
The Suspend Non-Lethal
function is enabled. Alarm
processing for numeric
physiological, technical and
non-lethal alarms has been
suspended for the selected tile.
Resume alarm processing by
selecting the Suspend
Non-Lethal sidebar button
and disabling the Suspend
Non-Lethal function.
These settings are
currently being edited
remotely.
The Alarm Limits Menu is
opened at the bedside monitor
while the Alarm Limits tab is
active at the central station
The Alarm Limits tab is active at
the central station and then the
Alarm Limits Menu is opened at
the bedside monitor.
The Print menu bar button is
disabled.
The message These settings
are currently being edited
remotely. is displayed.
Select the OK button to close
the message box and enable
the sidebar button.
Close the Alarm Limits menu at
the bedside device.
The Suspend All Alarms
and Suspend Non-Lethal
sidebar buttons are disabled.
The message These settings
are currently being edited
remotely. is displayed.
Select the OK button to close
the message box and enable
the sidebar button.
Close the Alarm Limits menu at
the bedside device.
The patient alarm limits were
cl
eared from the Alarm
Limits tab.
The patient size was changed
at the bedside monitor.
No solution required. Within
20 seconds, the alarm limits
from the bedside will be
displayed at the central station.
* Messages are shown in all bold text.