Panorama™ Operating Instructions 0070-10-0690-01 6 - 11
Patient Alarm Functions Alarm Limits Tab (Patient)
The Restore Previous
Settings sidebar button not
functioning.
The previous alarm limits were
cleared because the Alarm
Limits tab was open while the
patient size was changed at
the bedside monitor.
No solution required. Within
20 seconds, the alarm limits
from the bedside will be
displayed at the central station.
The Suspend All Alarms
sidebar button is not displayed.
The Suspend Alarms option in
the System Alarm Options tab
option is set to OFF or
Non-Lethal.
If this feature is needed,
contact the System
Administrator to change the
System Alarm Options.
The Suspend Non-Lethals
and Suspend All Alarms
sidebar buttons are not
displayed.
The Suspend Alarms option in
the System Alarm Options tab
option is set to OFF.
If this feature is needed,
contact the System
Administrator to change the
System Alarm Options.
The Alarm Limits tab is
disabled at the WorkStation.
The central station from where
the patient tile was networked
is configured as View Only.
No solution required.
The Alarm Responses tab is
disabled at the WorkStation.
The central station from where
the patient tile was networked
is configured as View Only.
No solution required.
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.