User Manual
Table Of Contents
- 08-0308 - Manual Cover.pdf
- 0070-10-0690-01 Panorama Operations Maual with Telepack 2.4 - CH16 REVISED2.pdf
- Foreword
- Warnings, Precautions and Notes
- Indications for Use
- Unpacking Information
- Symbols and Descriptions
- Technical Support
- 1.0 System Overview
- 1.1 Components of the Panorama System
- 1.2 Licensing Key
- 1.3 Supported Devices
- 1.3.1 Panorama Central Station
- 1.3.2 Panorama ViewStation and WorkStation
- 1.3.3 Panorama Telemetry Server
- 1.3.4 Panorama Wireless Transceiver
- 1.3.5 Panorama Gateway
- 1.3.6 Panorama Web Viewer
- 1.3.7 Passport 2® Bedside Monitor
- 1.3.8 Spectrum™ Bedside Monitor
- 1.3.9 Spectrum OR™ Bedside Monitor
- 1.3.10 Panorama Telepack
- 1.3.11 Network Printer
- 1.4 Troubleshooting
- 2.0 Getting Started
- 3.0 Alarm Behavior
- 4.0 Arrhythmia and ST Analysis
- 5.0 Patient Setup Functions
- 5.1 Demographics Tab
- 5.2 Transfer Tab
- 5.3 Discharge Tab
- 5.4 Display Tab
- 5.5 Standby Tab
- 5.6 Wave Gain Tab (Patient)
- 5.7 Print Setup Tab (Patient)
- 6.0 Patient Alarm Functions
- 7.0 Views
- 7.1 Bedside Tab
- 7.2 Trends Tab
- 7.3 Events Tab
- 7.4 Disclosure Tab
- 7.4.1 Accessing the Disclosure Tab
- 7.4.2 Disclosure Tab (Waveforms View)
- 7.4.2.1 Accessing the Waveforms View
- 7.4.2.2 Components of the Waveforms View
- 7.4.2.3 Navigating in the Waveforms View of the Disclosure Tab
- 7.4.2.4 ST Button
- 7.4.2.5 Comments Dialog Box (Optional)
- 7.4.2.6 Sidebar Buttons
- 7.4.2.7 Printing a Full Disclosure Zoom In Report from the Waveforms View
- 7.4.2.8 Troubleshooting
- 7.5 ST Tab
- 8.0 Report Functions
- 8.1 Patient Reports Tab
- 8.2 System Reports Tab
- 8.3 Print Status Tab
- 8.4 Additional Reports
- 9.0 System Setup Functions
- 9.1 Parameter Color Tab
- 9.2 Touch Screen Tab
- 9.3 Installation Setup Tab
- 9.4 Care Group Tab
- 9.5 Volume Tab
- 9.6 System Alarms Tab
- 9.7 Passwords Tab
- 9.8 Equipment Setup Tab
- 9.9 More Tab
- 9.10 Previous Tab
- 9.11 Date/Time Tab
- 9.12 Print Setup Tab (System)
- 9.13 Unit Priorities Tab
- 9.14 Unit Choices Tab
- 9.15 Wave Gain Tab (System)
- 9.16 Wireless Tab
- 10.0 Telepack (Optional)
- 11.0 User Maintenance
- 12.0 Parameter Specifications
- 12.1 Parameter Specifications
- 12.1.1 Heart Rate: HR
- 12.1.2 Respiration: Resp
- 12.1.3 NIBP
- 12.1.4 SpO2
- 12.1.5 Nonin SpO2
- 12.1.6 IBP: mmHg
- 12.1.7 CO2
- 12.1.8 O2
- 12.1.9 INSP/ET O2
- 12.1.10 N2O
- 12.1.11 Anesthetic Agents: Iso, Enf, Des, Sev, Hal, Agent
- 12.1.12 Premature Ventricular Contraction (PVC)
- 12.1.13 Temp: T1,T2, Delta T, and T Blood
- 12.1.14 ST
- 12.1.15 Cardiac Output
- 12.1.16 Cardiac Index
- 12.1.17 Pulmonary Artery Wedge Pressure (PAWP)
- 12.1.18 Battery Percentage
- 12.2 Invalid Data Display
- 12.1 Parameter Specifications
- 13.0 Alarms and Events
- 14.0 Network Tab (Optional)
- 14.1 Network Tab
- 14.1.1 Accessing the Network Tab
- 14.1.2 Sorting the Contents of the Network Tab List Box
- 14.1.3 Assigning Patient Tiles in the Network Tab
- 14.1.4 Removing a Patient Tile Assignment in the Network Tab
- 14.1.5 Removing All Patient Tile Assignments in the Network Tab
- 14.1.6 Refreshing the Network Tab List Box
- 14.1.7 Troubleshooting
- 14.1 Network Tab
- 15.0 Paging Tab (Optional)
- 15.1 Paging Password Tab
- 15.2 Paging Tab
- 15.2.1 Accessing the Paging Tab
- 15.2.2 Adding a Pager
- 15.2.3 Assigning a Patient Tile
- 15.2.4 Adding All Patient Tiles to a Pager
- 15.2.5 Removing a Patient Tile Assignment
- 15.2.6 Removing All Patient Tiles from a Pager
- 15.2.7 Editing a Pager
- 15.2.8 Removing a Pager
- 15.2.9 Resending a Failed Page
- 15.2.10 Generating a Paging Report
- 15.2.11 Troubleshooting
- 17.0 Glossary
Panorama™ Operating Instructions 0070-10-0690-01 7 - 11
Views Bedside Tab
7.1.4.6 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
either of the Bedside tab views.
MESSAGE/ISSUE* REASON SOLUTION
A waveform parameter
is not displayed.
If the waveform parameter
exists, it has not been
scrolled into view on the
display.
Use the navigation buttons to scroll to the
desired parameter waveform.
The waveform parameter is
not being monitored at the
bedside.
Connect the sensor(s) for the waveform
parameter to the patient.
NIBP/Digital Data area
empty
No equipment is assigned to
the selected tile.
Assign a monitoring device to the patient
tile.
Patient is in the “Panorama
Standby Only” mode
Remove patient from Panorama
Standby Only by selecting the
Resume Monitoring button in the
Standby tab.
Patient is in the “Bedside and
Panorama Standby Only”
mode
Remove patient from Bedside and
Panorama Standby Only by
pressing the STANDBY button on front
panel of the bedside monitoring device.
Selected tile is in the
Discharge state.
Retrieve discharged patient.
(for NIBP area only) The equipment assigned to
the selected tile does not
support the NIBP parameter.
Assign a monitoring device that supports
NIBP to the selected tile.
Up arrow/down arrow
button is not available
or is disabled
View 12
™
card is in use Use double up/down arrow button to
navigate
There are less than six
waveforms available.
Double up
arrow/down arrow
button is not available
View 12
™
card is not in use
First waveform position
not affected when
navigation up/down
arrow button is used
The first waveform is not
affected by the navigational
buttons.
Lead/All ECG
sidebar button is
unavailable
No equipment is assigned to
the selected tile.
Assign a monitoring device to the patient
tile.
Patient is in the “Panorama
Standby Only” mode.
Remove patient from Panorama
Standby Only by selecting the
Resume Monitoring button in the
Standby tab.
Patient is in the “Bedside and
Panorama Standby Only”
mode.
Remove patient from Bedside and
Panorama Standby Only by
pressing the STANDBY button on front
panel of the monitoring device.
Selected tile is in the
Communications Lost state or
Discharge state.
Verify connection to central stations
and/or retrieve discharged patient.