User Manual

Table Of Contents
Bedside Tab Views
7 - 12 0070-10-0690-01 Panorama™ Operating Instructions
Selected device is a Telepack
with a 3 or 5-wire lead set.
Selected device is a
Passport 2 or Spectrum with
a 3 or 5-wire lead set.
Bedside tab is in the All ECG
display or Freeze mode.
Select the All ECG sidebar button to
change to the Normal display mode.
Lead sidebar button is
disabled
The selected device is a:
Telepack device with a 3
or 5-wire lead
Telepack with Nonin
SpO
2
module attached
with a 3-wire lead set
No solution required.
The All ECG sidebar button
is latched.
Select the All ECG sidebar button to
enable the Lead sidebar button.
Buttons in Measure
dialog box are
unavailable
Interval is not selected. Select an interval.
A waveform parameter
is not displayed.
If the waveform parameter
exists, it has not been
scrolled into view on the
display.
Use the View More sidebar button to
scroll to the desired parameter
waveform.
The waveform parameter is
not being monitored at the
bedside.
Connect the sensor(s) for the waveform
parameter to the patient.
The Print menu bar
button is disabled.
The patient tile is in the
Bedside and Panorama
Standby mode.
If necessary, remove the patient tile from
this mode.
The patient tile is in the
Panorama Standby
Only mode.
If necessary, remove the patient tile from
this mode.
The patient tile is in the
Communications Lost
state.
Restore communications with the bedside
monitor.
The patient tile is in a
Waiting for Patient state.
Restore communications with the bedside
monitor.
MESSAGE/ISSUE* REASON SOLUTION