Panorama™ Operating Instructions 0070-10-0690-01 7 - 41
Views Disclosure Tab
7.4.1.1 Components of the Compressed View
Patient Demographic Field
The patient demographic field displays the last name and bed number of the selected patient,
if this data was entered.
Waveform Area
The waveform area of the Disclosure tab provides a compressed view of the selected
patient’s historical waveform data. A maximum of 10 minutes of waveform data can be
scrolled into the viewable area of the display. Each row of historical disclosure data in the
waveform area is displayed in one minute intervals. Each waveform is displayed in the
designated color of the parameter.
On initial entry to the Compressed view of the Disclosure tab, the ECG waveform
displayed is the first waveform shown in the main screen.
The waveforms displayed in the Disclosure tab are based on the sensors currently in use.
Any previously collected data is displayed, even if the parameter is not currently being
monitored. Newly stored waveform data does not immediately update the disclosure
waveform area. Selecting another control in the Disclosure tab causes the waveform area
to refresh and updates the newly stored waveform data.
Blank sections of a disclosure waveform occur during the storage of a row of data when:
The system time has been altered
The device has been placed into Standby mode
The device is in a Communications Lost state
The Panorama Central Station has been restarted
The entire waveform area will be blank if the selected patient tile does not have a monitoring
device attached or if there is no historical disclosure data in the patient’s database.
Time, Date, and Sensor Label
The actual time the data was stored is displayed on the left side of the waveform. The format
of the time display follows the System Time Format as defined in the ‘‘Date/Time Tab’’ on
page 9-41.
The date the selected waveform data was recorded is displayed in the upper right corner of
the waveform area (shown in FIGURE 7-20).
A sensor label for the selected waveform data is also displayed in the upper right corner of
the waveform area, just below the date (shown in FIGURE 7-20).
Event Markers
The Disclosure tab uses event markers to indicate that an event (such as an alarm
condition) has occurred. Event markers are small, color coded, horizontal lines that are
drawn above the disclosure waveform at the onset of the event. The color codes for the event
markers are defined as follows: