User Manual

Table Of Contents
Disclosure Tab Views
7 - 52 0070-10-0690-01 Panorama™ Operating Instructions
A section of the waveform
is blank.
The system time was altered
during the storage of the row
of data.
Wait for new data to accumulate.
The device was placed into
Standby mode during the
storage of the row of data.
Restore the system to normal
operating mode to resume data
collection.
The device experienced a
Communications Lost
event during the storage of the
row of data.
Determine the cause of the
Communications Lost condition
and rectify to resume data collection.
The Panorama was restarted
during the storage of the row
of data.
Wait for new data to accumulate.
No more events found
(Previous Event Button)
There is no event data in the
Disclosure database.
Wait until an event is recorded to the
Disclosure database.
There are no more events in
the patient’s Disclosure
database that occurred
before the time at the upper
left corner of the compressed
view.
Select the OK button to close the
message box and return to the
Disclosure tab.
The oldest data is at the upper
left corner of the compressed
view.
Select the OK button to close the
message box and return to the
Disclosure tab.
No more events found
(Next Event Button)
There is no event data in the
Disclosure database.
Wait until an event is recorded to the
Disclosure database.
There are no more events in
the patient’s Disclosure
database that occurred after
the time at the lower right
corner of the compressed
view.
Select the OK button to close the
message box and return to the
Disclosure tab.
The most recent data is at the
lower right corner of the
compressed view.
Select the OK button to close the
message box and return to the
Disclosure tab.
Time out of range
(Page Up button)
There are less than 10 rows of
data that are newer than the
top row.
Select the OK button to close the
message box and return to the
Disclosure tab. Wait for new data
to accumulate.
Time out of range
(Page Down button)
There are less than 10 rows of
data that are older than the
top row.
Select the OK
button to close the
mess
age box and return to the
Disclosure tab. Data is not
available.
Time out of range
(single up-arrow button)
There is no new data to
display.
Select the OK button to close the
message box and return to the
Disclosure tab. Wait for new data
to accumulate.
Time out of range
(single down-arrow button)
There is no older data to
display.
Select the OK button to close the
message box and return to the
Disclosure tab. Data is not
available.
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.