User Manual
Table Of Contents
- 08-0308 - Manual Cover.pdf
- 0070-10-0690-01 Panorama Operations Maual with Telepack 2.4 - CH16 REVISED2.pdf
- Foreword
- Warnings, Precautions and Notes
- Indications for Use
- Unpacking Information
- Symbols and Descriptions
- Technical Support
- 1.0 System Overview
- 1.1 Components of the Panorama System
- 1.2 Licensing Key
- 1.3 Supported Devices
- 1.3.1 Panorama Central Station
- 1.3.2 Panorama ViewStation and WorkStation
- 1.3.3 Panorama Telemetry Server
- 1.3.4 Panorama Wireless Transceiver
- 1.3.5 Panorama Gateway
- 1.3.6 Panorama Web Viewer
- 1.3.7 Passport 2® Bedside Monitor
- 1.3.8 Spectrum™ Bedside Monitor
- 1.3.9 Spectrum OR™ Bedside Monitor
- 1.3.10 Panorama Telepack
- 1.3.11 Network Printer
- 1.4 Troubleshooting
- 2.0 Getting Started
- 3.0 Alarm Behavior
- 4.0 Arrhythmia and ST Analysis
- 5.0 Patient Setup Functions
- 5.1 Demographics Tab
- 5.2 Transfer Tab
- 5.3 Discharge Tab
- 5.4 Display Tab
- 5.5 Standby Tab
- 5.6 Wave Gain Tab (Patient)
- 5.7 Print Setup Tab (Patient)
- 6.0 Patient Alarm Functions
- 7.0 Views
- 7.1 Bedside Tab
- 7.2 Trends Tab
- 7.3 Events Tab
- 7.4 Disclosure Tab
- 7.4.1 Accessing the Disclosure Tab
- 7.4.2 Disclosure Tab (Waveforms View)
- 7.4.2.1 Accessing the Waveforms View
- 7.4.2.2 Components of the Waveforms View
- 7.4.2.3 Navigating in the Waveforms View of the Disclosure Tab
- 7.4.2.4 ST Button
- 7.4.2.5 Comments Dialog Box (Optional)
- 7.4.2.6 Sidebar Buttons
- 7.4.2.7 Printing a Full Disclosure Zoom In Report from the Waveforms View
- 7.4.2.8 Troubleshooting
- 7.5 ST Tab
- 8.0 Report Functions
- 8.1 Patient Reports Tab
- 8.2 System Reports Tab
- 8.3 Print Status Tab
- 8.4 Additional Reports
- 9.0 System Setup Functions
- 9.1 Parameter Color Tab
- 9.2 Touch Screen Tab
- 9.3 Installation Setup Tab
- 9.4 Care Group Tab
- 9.5 Volume Tab
- 9.6 System Alarms Tab
- 9.7 Passwords Tab
- 9.8 Equipment Setup Tab
- 9.9 More Tab
- 9.10 Previous Tab
- 9.11 Date/Time Tab
- 9.12 Print Setup Tab (System)
- 9.13 Unit Priorities Tab
- 9.14 Unit Choices Tab
- 9.15 Wave Gain Tab (System)
- 9.16 Wireless Tab
- 10.0 Telepack (Optional)
- 11.0 User Maintenance
- 12.0 Parameter Specifications
- 12.1 Parameter Specifications
- 12.1.1 Heart Rate: HR
- 12.1.2 Respiration: Resp
- 12.1.3 NIBP
- 12.1.4 SpO2
- 12.1.5 Nonin SpO2
- 12.1.6 IBP: mmHg
- 12.1.7 CO2
- 12.1.8 O2
- 12.1.9 INSP/ET O2
- 12.1.10 N2O
- 12.1.11 Anesthetic Agents: Iso, Enf, Des, Sev, Hal, Agent
- 12.1.12 Premature Ventricular Contraction (PVC)
- 12.1.13 Temp: T1,T2, Delta T, and T Blood
- 12.1.14 ST
- 12.1.15 Cardiac Output
- 12.1.16 Cardiac Index
- 12.1.17 Pulmonary Artery Wedge Pressure (PAWP)
- 12.1.18 Battery Percentage
- 12.2 Invalid Data Display
- 12.1 Parameter Specifications
- 13.0 Alarms and Events
- 14.0 Network Tab (Optional)
- 14.1 Network Tab
- 14.1.1 Accessing the Network Tab
- 14.1.2 Sorting the Contents of the Network Tab List Box
- 14.1.3 Assigning Patient Tiles in the Network Tab
- 14.1.4 Removing a Patient Tile Assignment in the Network Tab
- 14.1.5 Removing All Patient Tile Assignments in the Network Tab
- 14.1.6 Refreshing the Network Tab List Box
- 14.1.7 Troubleshooting
- 14.1 Network Tab
- 15.0 Paging Tab (Optional)
- 15.1 Paging Password Tab
- 15.2 Paging Tab
- 15.2.1 Accessing the Paging Tab
- 15.2.2 Adding a Pager
- 15.2.3 Assigning a Patient Tile
- 15.2.4 Adding All Patient Tiles to a Pager
- 15.2.5 Removing a Patient Tile Assignment
- 15.2.6 Removing All Patient Tiles from a Pager
- 15.2.7 Editing a Pager
- 15.2.8 Removing a Pager
- 15.2.9 Resending a Failed Page
- 15.2.10 Generating a Paging Report
- 15.2.11 Troubleshooting
- 17.0 Glossary
Panorama™ Operating Instructions 0070-10-0690-01 7 - 53
Views Disclosure Tab
Time out of range
(Done button in the Skip
To dialog box)
A time and date that is older
than the oldest data or newer
than the most recent data was
entered in the Skip To dialog
box.
Select the OK button to close the
message box. Select the Skip To
sidebar button and enter a time and
date that is within the time frame of
the database.
ST button is disabled ST was never enabled for the
selected patient tile.
If the patient requires ST analysis,
and there is an ST license available,
enable ST for the selected patient
tile.
Comments sidebar
button is not displayed in
sidebar
The Paging option is not
present on the central station
licensing key.
Obtain a licensing key that contains
the Paging option.
There is a gap between
waveform times
Either the central station and
monitoring device were not
communicating during that
time (Communications Lost) or
the system time was changed
in the Date/Time tab.
No solution required.
Comments sidebar
button is disabled.
No paging information was
sent for the selected event.
If a page should have been sent for
the selected event type, verify the
Page On Alarm settings in the Alarm
Responses tab.
The current event was
deleted from the
database. Press the
OK button to refresh
the data.
This message will be
displayed when viewing a
patient tile’s Full Disclosure
Zoom In tab at the
ViewStation or WorkStation,
and the Next Event and/or
Previous Event button is
selected to view an event that
has been deleted from the
database.
Select the OK button to close the
message box.
The Previous Event
button is disabled.
The cursor is positioned at the
first record of the Disclosure
database.
No solution required.
Lead sidebar button is
disabled
The selected device is a:
• Telepack device with a 3
or 5-wire lead set
• Telepack with Nonin
SpO
2
module attached
with a 3-wire lead set
No solution required.
The All ECG sidebar button is
latched.
Select the All ECG sidebar button to
enable the Lead sidebar button.
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.