User Manual

Table Of Contents
ST Tab Views
7 - 56 0070-10-0690-01 Panorama™ Operating Instructions
No License indicates that an ST license is not available for the selected patient tile.
Disabled indicates that the ST algorithm on the bedside monitoring device is currently
disabled.
Relearning indicates that the ST algorithm is enabled and is in the learning phase of ST
analysis.
Real Time indicates that the ST algorithm is enabled, a successful learn phase has been
completed, and the most current data is being displayed.
Patient in Standby indicates that the selected patient tile is in a standby state.
Communications Lost indicates that the Panorama Central Station is not
communicating with the monitoring device.
ST Template Area (Device Dependent)
The ST template area will display ST templates and digital data for each ECG lead that is
utilized when using a hardwired bedside device, a wireless 2.4 GHz bedside device, or a
Telepack device. ECG lead labels are shown at the top of the ST template area. These labels
represent the ST data currently displayed.
NOTE: When using a 3 or 5-wire lead set, ST is invalidated when
the measured ST value exceeds range, and/or paced
rhythm persists for more than 45 seconds.
When using a View 12
card, ST is invalidated when the
measured ST value exceeds range, and/or paced rhythm
persists for more than 30 seconds.
For all lead wire sets, ST is also invalidated during detected
episodes of Asystole, V-Fib, V-Tach, and V-Rhythm.
NOTE: ST templates are not displayed when using a WMTS 608
bedside device.
When utilizing a 3-wire lead set, the ST tab will contain one ST template for the lead that
is currently being monitored.
When utilizing a 5-wire lead set, the ST tab will contain seven ST templates.
When utilizing a View 12
card, the ST tab will contain 12 ST templates.
Reference ST
The reference ST template is displayed in white
The digital data for the reference ST displays at the bottom of the ST template area
The reference ST template displays dashes (- - -) when ST is relearning, when there is no
data available, or when the data is invalid
Current ST
The current ST template is displayed in yellow
The digital data for the current ST will be shown at the bottom of the ST template area
The current ST template will display dashes (- - -) when ST is relearning, when there is no
data available, and when the data is invalid
NOTE: ST templates and digital data will not be shown in the ST
template areas during an ST learn phase.