User Manual

Table Of Contents
System Reports Tab Report Functions
8 - 32 0070-10-0690-01 Panorama™ Operating Instructions
8.2.2.2 Digital Data
The All Strips (All Patients) Report’s digital data section displays measurements for the digital
data parameters currently available for the selected patient tile. The displayed parameter
order is determined by the system Unit Priorities tab. The digital data section contains the
following:
A parameter label
A numeric measurement
A unit of measure label
A heart rate (HR) value
Dashes (--) to indicate missing or invalid digital data.
Square brackets ( [ ] ) to indicate alarming parameter values
NOTE: Digital Data is only shown on the first page of the report.
8.2.2.3 Waveform Data
The waveforms that print in the All Strips (All Patients) Report’s waveform data section are
limited to which leads are currently in use, and on the waveforms selected in the Print Setup
function.
The first page of this report contains up to two waveforms while subsequent pages contain up
to three waveforms. Each waveform in this report represents approximately 10 seconds of
data; five seconds prior to the time of the print request and 5 seconds after the print request.
A waveform label (i.e., ECG Lead II) is shown above each strip to identify the data source
and the unit of measurement (i.e.,%, RPM, BPM, mmHg, kPa). Pleth, ECG and Respiration do
not display a unit of measurement. The waveforms displayed in the report are shown on a
calibrated grid and contain a scale bar.
For pacemaker patients that have Pacer Enhancement enabled, the report displays a vertical
bar to indicate the firing of the pacemaker impulse.
ECG Waveforms
The ECG waveforms shown in the All Strips (All Patients) Report are based on the ECG lead
wire set currently in use and on the waveforms selected in the Print Setup tab. The ECG
waveform report displays with the following: