Panorama™ Operating Instructions 0070-10-0690-01 9 - 17
System Setup Functions System Alarms Tab
NOTE: Changes to the Suspend Alarms option will affect all
patients upon entering the Patient Alarms tab.
8. Select the ST Alarm Delay button to set an alarm delay for ST alarms. ST alarm
delay is the period of time that the Panorama Central Station will wait before ST alarm
notification. ST Alarm Delay options include 30 seconds, 45 seconds, 1 minute, 90
seconds, 2 minutes, and 3 minutes. The factory default setting is 30 seconds.
NOTE: ST Alarm Delay only affects new ST alarms for Telepack
devices. When using a bedside monitor, the ST alarm delay
settings are transferred to the central station.
9. Select the Alarm Delay button to set an alarm delay for the Panorama Central Station.
The alarm delay is the period of time that the Panorama Central Station will wait before
alarm notification. Alarm Delay options include None, 1 second, 2 seconds, 3 seconds,
4 seconds, 5 seconds, 6 seconds, 7 seconds, and 8 seconds. The factory default setting
is None.
NOTE: Alarm Delay only affects new alarms for Telepack devices.
When using a bedside monitor, the ST alarm delay settings
are transferred to the central station.
10. Select the Suspend All Timeout button to temporarily suspend all alarms or all
non-lethal alarms for a configured time period. This button is only enabled when the
Suspend Alarms option is set to All. When this option is enabled, and a time period
is selected, either the Suspend All Alarms or the Suspend Non-Lethal sidebar
button in the Patient Alarm Limits and Patient Alarm Responses tab must be selected
before the alarm suspension begins. Suspend All Timeout options include 2 minutes,
3 minutes, 5 minutes, 10 minutes, and None. The factory default setting is 2 minutes.
9.6.1.3 Setting the System Alarm Options (ViewStation and WorkStation)
This section outlines selecting system alarm option defaults at the ViewStation and the
WorkStation.
The ViewStation and the WorkStation are independent of each other. For example, if an
option is selected at the ViewStation, the WorkStation will not be affected unless the same
option is selected at the WorkStation.
1. Select the Password Protection toggle button to either enable or disable the
password protection for the Patient Alarms tab at the WorkStation. Password
Protection options include ON and OFF. The factory default setting is ON.
•The ON option indicates that Password Protection will be enabled when trying to
access the Patient Alarms tab.
•The OFF options indicates that Password Protection will not be enabled when
trying to access the Patient Alarms tab.
NOTE: The Password Protection toggle button is not available at
the ViewStation.
NOTE: Changes to the Password Protection option will affect all
patients upon entering the Patient Alarms tab at the
WorkStation.