Panorama™ Operating Instructions 0070-10-0690-01 1 - 11
System Overview Supported Devices
1.3.9 Spectrum OR
Bedside Monitor
Similar to the Spectrum, Spectrum OR bedside monitors are used to monitor patient data and
send it to the Panorama Central Station for viewing and storage via an Ethernet connection.
The Spectrum OR requires a hardwired configuration. The data collected by a Spectrum OR
is dependent on the type of lead wire set and sensors used.
NOTE: The CO/CI, Temp 2, and T Blood parameters are only
available if they are being monitored by the Spectrum/
Spectrum OR External Parameter Module (EPM). These
parameters are not available if they are monitored by an
Edwards Vigilance
®
monitor.
Parameters/Waveforms
PARAMETERS SETTINGS PANORAMA ALARMS
ECG/Heart Rate Leads I, II, III are available with a 3-wire
lead set.
Low Heart Rate
High Heart Rate
Leads I, II, III, aVR, aVL, aVF, V are
available with a 5-wire lead set.
ST ST analysis is in use at the bedside
device.
ST Single
ST Dual
PVC/min. Arrhythmia analysis is in use at the
bedside device.
Asystole
Bigeminy
Brady
2
Couplet
Irregular Heart Rate
2
Pause
1
High PVC
Run
Trigeminy
V-Fib
V-Rhythm
V-Tach
Respiration (Source:
ECG, CO
2
and Gas
Module)
Respiration is in use at the bedside and
is derived from a valid source.
Low Respiration
High Respiration
Apnea
SpO
2
SpO
2
sensor is in use at the bedside
device.
Low SpO
2
High SpO
2
IBP1/IBP2/IBP3/IBP4
(Systolic, Diastolic,
Mean)
IBPx option is in use at the bedside
device.
IBP parameter labels include IBP, Art,
UA, LV, PA, CVP, ICP, LA and RA. The
label used is based on the label selected
at the bedside monitor.
Low IBPx Systolic
High IBPx Systolic
Low IBPx Diastolic
High IBPx Diastolic
Low IBPx Mean
High IBPx Mean
NOTE: x = 1, 2, 3 or 4 (IBP3 and 4 are only available if the
Spectrum/Spectrum OR External Parameter Module
(EPM) is installed)
INSP/ET CO
2
Gas Module is in use at the bedside
device.
High CO
2
Inspired
Low CO
2
End Tidal
High CO
2
End Tidal Rate
INSP/ET O
2
Gas Module is in use at the bedside
device.
O
2
Inspired
O
2
End Tidal
INSP/ET N
2
O Gas Module is in use at the bedside
device.
N
2
O Inspired
N
2
O End Tidal