9 - 52 0070-10-0690-01 Panorama™ Operating Instructions
3. Select the Pacer Reject button to select the system default pacer reject period. Pacer
Reject adjusts the time period at which pacemaker-induced ECG complexes are filtered
for ECG sampling processes.
Pacer Reject options include 0 to 80 milliseconds (ms) in 20-second increments. The
factory default setting is 40 ms.
NOTE: Changing the Pacer Reject option will affect all newly
admitted patients and will have no affect on patients
currently being monitored.
NOTE: The Pacer Reject option is not available at the ViewStation
or the WorkStation.
4. Select the Paging Demographics button to define which demographic information is
sent to the paging system when a page is requested. Paging Demographics options
include Bed, Bed + Last Name, and Bed + ID. The factory default setting is Bed.
Refer to the following tables for formatting examples:
NOTE: The amount of data that is displayed in the pager display
window may differ slightly due to the type of pager being
used.
NOTE: Changing the Paging Demographics option will take effect
immediately.
OPTION ALARM TYPE FORMAT EXAMPLES
Bed Heart Rate central name:XXXBed:ICU10 Low Heart Rate Alarm
Bed Arrhythmia central name:XXXBed:ICU10 Asystole
Bed Not HR or
arrhythmia alarm
central name:XXXBed:ICU10 High Temperature 1
Bed ST alarm central name:XXXBed:ICU10 ST Single Alarm
OPTION ALARM TYPE FORMAT EXAMPLES
Bed + Last
Name
Heart Rate central name:XXXBed:ICU10 Name:Patient Last Name
Low Heart Rate Alarm
Bed + Last
Name
Arrhythmia central name:XXXBed:ICU10 Name:Patient Last Name
Asystole
Bed + Last
Name
Not HR or
arrhythmia alarm
central name:XXXBed:ICU10 Name:Patient Last Name
High Temperature 1
Bed + Last
Name
ST alarm central name:XXXBed:ICU10 Name:Patient Last Name ST
Single Alarm
OPTION ALARM TYPE FORMAT EXAMPLES
Bed + ID Heart Rate central name:XXXBed:ICU10 ID:Patient ID Low Heart Rate
Alarm
Bed + ID Arrhythmia central name:XXXBed:ICU10 ID:Patient ID Asystole
Bed + ID Not HR or
arrhythmia alarm
central name:XXXBed:ICU10 ID:Patient ID High
Temperature 1
Bed + ID ST alarm central name:XXXBed:ICU10 ID:Patien
t ID ST Single
Alarm