2. Use the dial control to change the system default V-Tach Rate. The range available is
100-180 bpm, in 5 bpm increments. The factory default setting is 120 bpm.
3. Select the Asystole Delay button to change the system default asystole delay time for
Arrhythmia processing. The range available is 3-10 seconds. The factory default setting
is 4 seconds.
4. Select the V-Tach Threshold button to change the system default V-Tach threshold
settings. The range available is 3-15 beats. The factory default setting is 3 beats.
5. Select the ECG Noise Delay button to change the system default noise delay setting.
The range available is 3-30 seconds. The factory default setting is 6 seconds.
NOTE: The ECG Noise Delay button settings only affect Telepack
monitoring devices.
NOTE: When using a bedside device, these settings are taken from
the bedside upon admission.
6. Accept or reject the selections made in the dialog box.
Select the Done button to accept the selections.
Select the Cancel button to discard the selections, close the dialog box and return to the
Unit Choices tab.
9.14.2.4 Settings (Telepack)
Select the Settings sidebar button to establish the default ECG monitoring filter that will be
used by the Telepack, and to select the ECG leads for ST monitoring.
NOTE: The Settings sidebar button is not available at the
ViewStation or the WorkStation.
The Monitor mode for ECG filtering establishes the frequency response range for non-ST
ECG monitoring.
The ST mode for ECG filtering establishes the frequency response range for ST diagnostic
monitoring.
NOTE: The Telepack can only have one type of ECG filtering
enabled at any given time (Monitor mode or ST mode).
To configure the default ECG monitoring filter:
1. Select the Settings sidebar button in the Unit Choices tab. The Settings dialog box
(shown in FIGURE 9-42) is displayed.
FIGURE 9-42 Settings Dialog Box