Panorama™ Operating Instructions 0070-10-0690-01 9 - 57
System Setup Functions Unit Choices Tab
2. Select the ST check box to enable or disable ST analysis for the selected patient.
A checkmark in the ST box indicates that ST analysis is enabled.
An empty box indicates that ST analysis is disabled.
When the ST check box is empty, the Monitor filter is automatically enabled.
NOTE: If ST licenses are available, an ST license is automatically
assigned to the patient when ST is enabled
.
NOTE: When using a 5-wire lead set or a View 12
card, ST is
performed on the leads available.
When using a 3-wire lead set, ST is performed on the lead
being viewed.
3. Select the Monitor or ST filter mode.
NOTE: One filter box must always be populated. When a filter
mode is selected, the other mode is automatically disabled.
A checkmark in the Monitor filter box indicates that the Monitor filter is enabled and
ST analysis is disabled.
A checkmark in the ST filter box indicates that the ST analysis is enabled and the
Monitor filter is disabled.
4. Accept or reject the selections made in the dialog box.
Select the Done button to accept the selections.
Select the Cancel button to discard the selections, close the dialog box and return to
the Unit Choices tab.
9.14.3 Configuring Default Report Waveform Parameters
Select the Strip Setup sidebar button (shown in FIGURE 9-38) to select the default
waveform parameters that will print in the All Strips, Event Zoom-In, and Full Disclosure
Zoom-In reports.
NOTE: The Strip Setup sidebar button is not available at the
ViewStation or the WorkStation.
To select the default waveform parameters:
1. From the Unit Choices tab, select the Strip Setup sidebar button.
FIGURE 9-43 Strip Setup View of the Unit Choices Tab