Panorama™ Operating Instructions 0070-10-0690-01 13 - 3
Alarms and Events Physiological Alarms and Events
ALARM
ALARM
PRIORITY ALARM LIMIT NEONATE PEDIATRIC ADULT DEVICES
Heart Rate Units: bpm 1 or 2
Factory Default: 2
High Limit Off, 100-350
Factory Default: 200
Off, 100-300
Factory Default:180
Off, 60-250
Factory Default:150
Passport 2
®
Spectrum
/Spectrum OR
Telepack
Low Limit Off, 30-200
Factory Default: 100
Off, 30-150
Factory Default: 80
Off, 30-120
Factory Default: 50
Heart Rate Units: bpm 1 or 2
Factory Default: 2
High Limit N/A N/A Off, 60-250
Factory Default:150
Telepack with Nonin SpO
2
module
Low Limit N/A N/A Off, 30-120
Factory Default: 50
NOTE: When the central station language is configured for French, the
Heart Rate (HR) high/low alarm limit cannot be turned Off.
NOTE: When the selected device is a Telepack, the Low Heart Rate alarm
will be called when the heart rate becomes invalid because the rate
is underrange.
Respiration
Units: rpm
1, 2, and 3
Factory Default: 3
High Limit Off, 30-200
Factory Default: Off
Off, 15-150
Factory Default: Off
Off, 10-100
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit Off, 5-50
Factory Default: 5
Off, 5-40
Factory Default: Off
Off, 5-30
Factory Default: Off
SpO
2
Units: %
1 and 2
Factory Default: 2
High Limit Off, 80-100%
Factory Default: Off
Off, 80-100%
Factory Default: Off
Off, 80-100%
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit 50-99%
Factory Default: 92%
50-99%
Factory Default: 90%
50-99%
Factory Default: 85%
SpO
2
(Nonin)
Units: %
1 and 2
Factory Default: 2
High Limit N/A N/A Off, 80-100%
Factory Default: Off
Telepack with Nonin SpO
2
module
Low Limit N/A N/A 50-99%
Factory Default: 85%
NOTE: The SpO
2
high/low alarm limit will always be displayed in the
digital tile portion of the Bedside tab All Waveforms view.
IBPx Sys.
Units: mmHg*
2 or 3
Factory Default: 3
High Limit Off, 5-180
Factory Default: Off
Off, 5-240
Factory Default: Off
Off, 5-300
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit Off, 0-130
Factory Default: Off
Off, 0-130
Factory Default: Off
Off, 0-150
Factory Default: Off
N/A - Not Applicable