Panorama™ Operating Instructions 0070-10-0690-01 13 - 5
Alarms and Events Physiological Alarms and Events
ET CO
2
(mmHg) 2 or 3
Factory Default: 3
High Limit Off, 20-80
Factory Default: 60
Off, 20-80
Factory Default: 60
Off, 20-80
Factory Default: 60
Passport 2
®
Spectrum
/Spectrum OR
Low Limit Off, 5-50
Factory Default: Off
Off, 5- 0
Factory Default: Off
Off, 5-50
Factory Default: Off
ET CO
2
(%) 2 or 3
Factory Default: 3
High Limit Off, 2-10
Factory Default: 8
Off, 2-10
Factory Default: 8
Off, 2-10
Factory Default: 8
Passport 2
Spectrum/Spectrum OR
Low Limit Off, 1-6
Factory Default: Off
Off, 1-6
Factory Default: Off
Off, 1-6
Factory Default: Off
ET CO
2
(kPa) 2 or 3
Factory Default: 3
High Limit Off, 2-10
Factory Default: 8
Off, 2-10
Factory Default: 8
Off, 2-10
Factory Default: 8
Passport 2
Spectrum/Spectrum OR
Low Limit Off, 1-6
Factory Default: Off
Off, 1-6
Factory Default: Off
Off, 1-6
Factory Default: Off
T1/T2
Units: F°
2 or 3
Factory Default: 3
High Limit Off, 95-110°
Factory Default: Off
Off, 95-110°
Factory Default: Off
Off, 95-110°
Factory Default: Off
T1 Passport 2
and Spectrum/
Spectrum OR
T2 optional for Spectrum/
Spectrum OR
Low Limit Off, 80-100°
Factory Default: Off
Off, 80-100°
Factory Default: Off
Off, 80-100°
Factory Default: Off
T1/T2
Units: C°
2 or 3
Factory Default: 3
High Limit Off, 35-43°
Factory Default: Off
Off, 35-43°
Factory Default: Off
Off, 35-43°
Factory Default: Off
T1 Passport 2
and Spectrum/
Spectrum OR
T2 optional for Spectrum/
Spectrum OR
Low Limit Off, 26-38°
Factory Default: Off
Off, 26-38°
Factory Default: Off
Off, 26-38°
Factory Default: Off
Delta T
Units: F°
2 or 3
Factory Default: 3
High Limit Off, 2-10°
Factory Default: Off
Off, 2-10°
Factory Default: Off
Off, 2-10°
Factory Default: Off
Spectrum/Spectrum OR
Low Limit Off, 2-10°
Factory Default: Off
Off, 2-10°
Factory Default: Off
Off, 2-10°
Factory Default: Off
Delta T
Units: C°
2 or 3
Factory Default: 3
High Limit Off, 1-5°
Factory Default: Off
Off, 1-5°
Factory Default: Off
Off, 1-5°
Factory Default: Off
Spectrum/Spectrum OR
Low Limit Off, 1-5°
Factory Default: Off
Off, 1-5°
Factory Default: Off
Off, 1-5°
Factory Default: Off
NOTE: Only available if Spectrum/Spectrum OR External Parameter Module (EPM) is installed and sensor is
attached
ALARM
ALARM
PRIORITY ALARM LIMIT NEONATE PEDIATRIC ADULT DEVICES
N/A - Not Applicable