User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 13 - 11
Alarms and Events Technical Alarms and Events
13.3 Technical Alarms and Events
The Panorama Central Station uses Technical alarms to indicate that a patient specific
technical issue has occurred.
This section provides the technical event messages that may be posted to the Events database
and shown in white text (with the exception of the All Alarms Suspended and Non-Lethal
Alarms Suspended events) in the Patient Status line. When a number is indicated beside the
event message, refer to the table footnotes.
Technical Events
MESSAGE REASON
All Alarms Suspended
1,2
The Suspend All Alarms sidebar button was
selected in the Patient Alarms tabs.
NOTE: This event message will display
in red text in the Patient Status
line.
Resume All Alarms
1,2,3
A patient’s alarms are to resume.
Non-Lethal Alarms Suspended
1,2
The Suspend Non-Lethal sidebar button is
selected in the Patient Alarms tabs.
NOTE: This event message will display
in yellow text in the Patient
Status line.
Resume Non-Lethal Alarms
1,2,3
A patient’s non-lethal alarms are set to resume.
Communications Lost
4,5
The connection between the Panorama Central
Station and the attached monitoring equipment
has been interrupted.
Central Communications Lost
4
The connection between the host central station
and the ViewStation has been interrupted.
Signal Lost-Limited Arrhythmia
4
A Telepack is in use, arrhythmia analysis is
enabled, and there are multiple disruptions of
data that may affect the performance of the
arrhythmia algorithm.
Signal Lost
4
A Telepack is in use, arrhythmia analysis is
disabled, and there are multiple disruptions of
data.
Standby
1,2
A patient was placed into Standby mode. In
Standby mode, all of the patient’s monitoring
activity is suspended.
NOTE: Message will not be shown in
the Patient Status line. Instead,
message and/or Standby
location will be shown in the
patient’s digital data tile in pink
text.
Resume Monitoring
1,2
The patient was removed from Standby mode.
1 Event cannot be deleted from Event database
2 Event will not elicit technical event sound
3 Event will not display in Patient Status line
4 Event will always elicit the technical event sound regardless of the Technical Event sound setting
5 Event will post a Page request when the central station has the paging option enabled and the alarm is
configured to send a page