Panorama™ Operating Instructions 0070-10-0690-01 13 - 13
Alarms and Events Technical Alarms and Events
LL Lead Off
4,5
The LL lead is not connected to the patient, ECG
lead II or III is selected while using a 3-wire lead,
5-wire lead set, or View 12
card.
Leads Off
4,5
Two or more of the ECG leads are not connected
to the patient and a 3-wire lead set, 5-wire lead
set or a View 12
card is in use. It will also be
shown when using a 3-wire lead set and ECG I,
II, or III is shown and the LL, LA, or RA leads are
off, respectively. When using a View 12
card
and two or more leads are off, unless one of the
disconnected leads is the RL lead. If a 5-wire lead
set is in use and the RL lead is off.
V Lead Off
4,5
A V lead is not connected to the patient and a
5-wire lead set is in use.
V1-V6 Lead Off
4,5
A V1-6 lead is not connected to the patient and a
View 12
card is in use.
ECG Noise
4
Noise is present on one or more leads.
RESPIRATION (ECG THORACIC IMPEDANCE) TECHNICAL EVENTS
Respiration CVA Present A CVA condition is detected for the monitored
patient.
SPO
2
(MASIMO SET
®
AND NELLCOR
®
) TECHNICAL EVENTS
SpO
2
System Check There is an SpO
2
system check.
SpO
2
Communication Error There is interference in the communication
between the SpO
2
source and the bedside
monitor.
SpO
2
Failure A faulty 5V power supply or a faulty positive or
negative 15V power supply is detected on the
SpO
2
board.
SpO
2
Board Fault The SpO
2
board fails.
SpO
2
Unrecognized Cable The system detects an unrecognized SpO
2
cable.
SpO
2
Pulse Search The SpO
2
sensor is attempting to detect a pulse.
SpO
2
Weak Pulse The SpO
2
sensor detects a weak pulse.
SPO
2
(MASIMO) TECHNICAL EVENTS
SpO
2
Low Perfusion The SpO
2
Masimo sensor detects low perfusion.
SpO
2
Too Much Light The SpO
2
Masimo sensor detects too much light.
SpO
2
Unrecognized Sensor An unrecognized SpO
2
Masimo sensor is
detected by the system.
SpO
2
Sensor Off The SpO
2
Masimo sensor is off.
Technical Events (Continued)
MESSAGE REASON
1 Event cannot be deleted from Event database
2 Event will not elicit technical event sound
3 Event will not display in Patient Status line
4 Event will always elicit the technical event sound regardless of the Technical Event sound setting
5 Event will post a Page request when the central station has the paging option enabled and the alarm is
configured to send a page