Panorama™ Operating Instructions 0070-10-0690-01 13 - 17
Alarms and Events System Events
System Service
required - Contact
Datascope or
Authorized Datascope
Distributor (100)
This message will be displayed after the central station and/or
workstation automatically restarted because a system issue was
detected.
System Service
required - Contact
Datascope or
Authorized Datascope
Distributor (300)
The central station or remote workstation has detected a hard drive
error.
CAUTION: It is important to notify Datascope Service
personnel when the system event message
“System Service required - Contact
Datascope or Authorized Datascope
Distributor (300).” is displayed. This
message indicates that while data was
being written to or read from the system
hard drive, an error was detected. Until
service personnel correct this issue, patient
data may be lost (for example,
demographic information, alarm
configurations, and historical patient
data). Real time parameter monitoring
and alarm checking will not be affected by
this message. This message will be
displayed until reset by service personnel.
System Event Messages
MESSAGE REASON