User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-07XX-02 16 - 13
Performance Specifications Panorama Central Station and Panorama Telemetry Server
TABLE 16-4
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE PANORAMA CENTRAL STATION AND
TELEMETRY SERVER
The Panorama Central Station, Telemetry Server, and Panorama Wireless Transceiver is intended for
use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Panorama Central Station, Telemetry Server, and Panorama Wireless
Transceiver can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Panorama Central
Station, Telemetry Server, and Panorama Wireless Transceiver as recommended below, according to
the maximum output power of the communications equipment.
RATED MAXIMUM
OUTPUT POWER OF
TRANSMITTER W (WATTS)
SEPARATION DISTANCE ACCORDING TO FREQUENCY
OF TRANSMITTER M (METERS)
150 KHZ TO
80 MHZ
D = 1.2 x
80 MHZ TO
800 MHZ
D = 1.2 x
800 MHZ TO
2.5 GHZ
D = 2.3 x
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
P
P
P