User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-07XX-02 16 - 25
Performance Specifications Telepack Specifications
TABLE 16-11
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE TELEPACK
The Telepack is intended for use in an electromagnetic environment where radiated RF disturbances
are controlled. Prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Telepack as recommended
in the following table, according to the maximum output power of the communications equipment.
RATED MAXIMUM
OUTPUT POWER OF
TRANSMITTER W (WATTS)
SEPARATION DISTANCE ACCORDING TO FREQUENCY
OF TRANSMITTER M (METERS)
150 KHZ TO
80 MHZ
D = 1.2 x
80 MHZ TO
800 MHZ
D = 1.2 x
800 MHZ TO
2.5 GHZ
D = 2.3 x
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE: The separation distances guidelines in Table 16-11 on
page 16-25 may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
P
P
P