User Manual
Table Of Contents
- 08-0308 - Manual Cover.pdf
- 0070-10-0690-01 Panorama Operations Maual with Telepack 2.4 - CH16 REVISED2.pdf
- Foreword
- Warnings, Precautions and Notes
- Indications for Use
- Unpacking Information
- Symbols and Descriptions
- Technical Support
- 1.0 System Overview
- 1.1 Components of the Panorama System
- 1.2 Licensing Key
- 1.3 Supported Devices
- 1.3.1 Panorama Central Station
- 1.3.2 Panorama ViewStation and WorkStation
- 1.3.3 Panorama Telemetry Server
- 1.3.4 Panorama Wireless Transceiver
- 1.3.5 Panorama Gateway
- 1.3.6 Panorama Web Viewer
- 1.3.7 Passport 2® Bedside Monitor
- 1.3.8 Spectrum™ Bedside Monitor
- 1.3.9 Spectrum OR™ Bedside Monitor
- 1.3.10 Panorama Telepack
- 1.3.11 Network Printer
- 1.4 Troubleshooting
- 2.0 Getting Started
- 3.0 Alarm Behavior
- 4.0 Arrhythmia and ST Analysis
- 5.0 Patient Setup Functions
- 5.1 Demographics Tab
- 5.2 Transfer Tab
- 5.3 Discharge Tab
- 5.4 Display Tab
- 5.5 Standby Tab
- 5.6 Wave Gain Tab (Patient)
- 5.7 Print Setup Tab (Patient)
- 6.0 Patient Alarm Functions
- 7.0 Views
- 7.1 Bedside Tab
- 7.2 Trends Tab
- 7.3 Events Tab
- 7.4 Disclosure Tab
- 7.4.1 Accessing the Disclosure Tab
- 7.4.2 Disclosure Tab (Waveforms View)
- 7.4.2.1 Accessing the Waveforms View
- 7.4.2.2 Components of the Waveforms View
- 7.4.2.3 Navigating in the Waveforms View of the Disclosure Tab
- 7.4.2.4 ST Button
- 7.4.2.5 Comments Dialog Box (Optional)
- 7.4.2.6 Sidebar Buttons
- 7.4.2.7 Printing a Full Disclosure Zoom In Report from the Waveforms View
- 7.4.2.8 Troubleshooting
- 7.5 ST Tab
- 8.0 Report Functions
- 8.1 Patient Reports Tab
- 8.2 System Reports Tab
- 8.3 Print Status Tab
- 8.4 Additional Reports
- 9.0 System Setup Functions
- 9.1 Parameter Color Tab
- 9.2 Touch Screen Tab
- 9.3 Installation Setup Tab
- 9.4 Care Group Tab
- 9.5 Volume Tab
- 9.6 System Alarms Tab
- 9.7 Passwords Tab
- 9.8 Equipment Setup Tab
- 9.9 More Tab
- 9.10 Previous Tab
- 9.11 Date/Time Tab
- 9.12 Print Setup Tab (System)
- 9.13 Unit Priorities Tab
- 9.14 Unit Choices Tab
- 9.15 Wave Gain Tab (System)
- 9.16 Wireless Tab
- 10.0 Telepack (Optional)
- 11.0 User Maintenance
- 12.0 Parameter Specifications
- 12.1 Parameter Specifications
- 12.1.1 Heart Rate: HR
- 12.1.2 Respiration: Resp
- 12.1.3 NIBP
- 12.1.4 SpO2
- 12.1.5 Nonin SpO2
- 12.1.6 IBP: mmHg
- 12.1.7 CO2
- 12.1.8 O2
- 12.1.9 INSP/ET O2
- 12.1.10 N2O
- 12.1.11 Anesthetic Agents: Iso, Enf, Des, Sev, Hal, Agent
- 12.1.12 Premature Ventricular Contraction (PVC)
- 12.1.13 Temp: T1,T2, Delta T, and T Blood
- 12.1.14 ST
- 12.1.15 Cardiac Output
- 12.1.16 Cardiac Index
- 12.1.17 Pulmonary Artery Wedge Pressure (PAWP)
- 12.1.18 Battery Percentage
- 12.2 Invalid Data Display
- 12.1 Parameter Specifications
- 13.0 Alarms and Events
- 14.0 Network Tab (Optional)
- 14.1 Network Tab
- 14.1.1 Accessing the Network Tab
- 14.1.2 Sorting the Contents of the Network Tab List Box
- 14.1.3 Assigning Patient Tiles in the Network Tab
- 14.1.4 Removing a Patient Tile Assignment in the Network Tab
- 14.1.5 Removing All Patient Tile Assignments in the Network Tab
- 14.1.6 Refreshing the Network Tab List Box
- 14.1.7 Troubleshooting
- 14.1 Network Tab
- 15.0 Paging Tab (Optional)
- 15.1 Paging Password Tab
- 15.2 Paging Tab
- 15.2.1 Accessing the Paging Tab
- 15.2.2 Adding a Pager
- 15.2.3 Assigning a Patient Tile
- 15.2.4 Adding All Patient Tiles to a Pager
- 15.2.5 Removing a Patient Tile Assignment
- 15.2.6 Removing All Patient Tiles from a Pager
- 15.2.7 Editing a Pager
- 15.2.8 Removing a Pager
- 15.2.9 Resending a Failed Page
- 15.2.10 Generating a Paging Report
- 15.2.11 Troubleshooting
- 17.0 Glossary
Arrhythmia Alarms Arrhythmia and ST Analysis
4 - 14 0070-10-0690-01 Panorama™ Operating Instructions
NOTE: Arrhythmia alarms are not available for the Neonate
patient size.
4.4.1 Lethal Arrhythmia Alarms
A lethal arrhythmia is an arrhythmia that can be life threatening to a patient if left untreated.
Ventricular Tachycardia (V-Tach), Ventricular Fibrillation (V-Fib), and Asystole alarms are
classified as lethal arrhythmia alarms. These alarms automatically default to Alarm Priority 1.
NOTE: Lethal arrhythmia alarms may be configured to have a
different alarm sound than other physiological alarms. The
sounds used are determined during system setup and
configuration.
NOTE: Lethal arrhythmia alarms are latched alarms and must be
acknowledged after the condition is resolved. To
acknowledge the alarm, select the digital data area or the
waveform area of the affected patient tile.
NOTE: If a lethal arrhythmia alarm condition exists, and a
transition to a different lethal arrhythmia alarm condition
occurs, a new event will be stored and a printout will be
generated, as configured.
NOTE: When a V-Tach or V-Fib alarm occurs and is replaced by an
Asystole alarm, the Asystole alarm will be paged. This will
occur if Paging has been installed on the central station; the
Asystole alarm paging delay timer has expired; and the
alarm condition still exists when the paging delay timer has
expired.
Asystole Alarm
An Asystole alarm is activated at the Panorama Central Station when no QRS complexes
are detected for the configured time period in the absence of Ventricular Fibrillation. When
using a View 12
™
card, the time period range for an Asystole alarm is between three (3)
and eight (8) seconds. When using a 3 or 5-wire lead set, the time period range for an
Asystole alarm is between three (3) and ten (10) seconds.
The Asystole alarm is a Priority 1 alarm event that:
• Produces an Alarm Priority 1 visual and audio alarm indicators once the condition is
identified.
• Causes a red Asystole text message to display in the Patient Status line.
Ventricular-Fibrillation (V-Fib) Alarm
A V-Fib alarm is activated at the Panorama Central Station when a fibrillated waveform (P,
QRS or T waves can no longer be identified) is detected. V-Fib is defined as “irregular,
disorganized electrical activity of the heart”. The ECG signal for V-Fib continuously changes
shape, interval, and amplitude. The V-Fib detection algorithm runs in parallel to the beat
detection algorithm and continuously examines the incoming data.
The V-Fib alarm is a Priority 1 alarm event that:
• Produces Alarm Priority 1 visual and audio alarm indicators once the condition is
identified.