Troubleshooting Arrhythmia and ST Analysis
4 - 20 0070-10-0690-01 Panorama™ Operating Instructions
No ECG Waveform Gain set too low. Readjust the ECG wave gain as
required. Refer to the Wave Gain
Tab (System) for instructions on
changing the wave gain setting.
NOTE: Changing the ECG
wave gain does not
affect the arrhythmia
algorithm.
Lead wires and patient cable
not fully or properly inserted.
Check for proper insertion.
Cable or lead wires
damaged.
Check with lead continuity tester.
Base Line Wander Patient moving excessively. Secure lead wires and cable to
patient.
Patient's respiration Reposition electrodes
Electrodes dry or loose Repeat skin preparation and apply
fresh, moist electrodes.
Static build up around patient. Check with biomedical engineer.
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.