3 - 2 TM80/TM70 Telemetry Monitor Operator’s Manual
3.1 Unpacking and Checking
Before unpacking, examine the packing case carefully for signs of damage. If any dam-
age is detected, contact the carrier or Mindray.
If the packing case is intact, open the package and remove the device and accessories
carefully. Check all materials against the packing list and check for any mechanical dam-
age. Contact Mindray in case of any problem.
WARNING
The network shall be installed by Mindray authorized personnel.
The software equipment copyright is solely owned by Mindray. No orga-
nization or individual shall resort to altering, copying, or exchanging it
or to any other infringement on it in any form or by any means without
due permission.
Connect only approved devices to this system. Devices connected to the
equipment must meet the requirements of the applicable IEC standards
(e.g. IEC 60950 safety standards for information technology equipment
and IEC 60601-1 safety standards for medical electrical equipment). The
system configuration must meet the requirements of the IEC 60601-1
medical electrical systems standard. Any personnel who connect
devices to the equipment’s signal input/output port is responsible for
providing evidence that the safety certification of the devices has been
performed in accordance to the IEC 60601-1. If you have any questions,
please contact Mindray.
If it is not evident from the equipment specifications whether a particu-
lar combination with other devices is hazardous, for example, due to
summation of leakage currents, please consult the manufacturers or
else an expert in the field, to ensure the necessary safety of patients and
all devices concerned will not be impaired by the proposed combina-
tion.
Contact Mindray to relocate the TM80/TM70.
Only Mindray authorized personnel can update the TM80/TM70.