c.  STEP 3  Test Piece 
•  Ensure  the  operator’s  hands  and  arms  are  free  of  metallic 
objects  (watches,  rings  etc),  and  that  no  other  metallic  objects 
are  near  the  coil.   The orientation of  the Test  Piece during  the 
test  is  dependent  on  which  Sensitivity  Endcap  is  connected  to 
the detector. 
NOTE 
Maximum sensitivity is only available 30 seconds after the Threshold 
Tone commences.  Do not test the detector with the Test Piece until 
30 seconds after the Threshold Tone begins. 
•  Black Sensitivity Endcap: 
Hold  the  Test  Piece  above  the  middle  of  the  coil  with  the 
rounded end (containing metallic target) AWAY from the coil. 
Move the Test Piece towards the centre of the coil until it lightly 
touches the surface then move it sideways off the coil (the Test 
Piece  should  be  moved  slowly  and  smoothly  during  this 
procedure). A faint but clear response (change in Threshold 
Tone volume and pitch) should be heard indicating the sensitivity 
of the detector is correct. 
P0647-A
Figure 20: Test Piece Procedure 