Instruction manual
82Understanding your X-TERRA - Technology
Q
A
 
In sweeping the 3 kHz low frequency (LF) and 7.5 kHz medium 
frequency (MF) 9-inch Concentric coils over an old U.S. silver 
dollar (90% silver), the TID displays a value of 48 at distances 
less than 2-inches from the coil. At distances greater than 
2-inches, the TID is 46. When using 18.75 kHz high frequcey 
(HF) Double-D coils, (6-inch, elliptical and 10.5-inch round) the 
same U.S. silver dollar registers a 48 regardless of the distance 
between the target and the coil. What causes this TID variation 
between coils?
This is a somewhat surprising eect, because one would expect 
the TID from MF and certainly LF to be more stable for a high 
conductivity target than that from the HF coil. Therefore, the 
question requires discussion on two fronts before I can oer an 
answer:
How TID’s are aected by coil frequency
As mentioned in the previous answer, the TID is calculated 
based on the phase angle of the received signal with respect 
to the transmitted signal. Each TID corresponds to a range of 
phase angles (a “bin”). For example, the X-TERRA 705 has 24 
non-ferrous TID’s in 90 degrees, so about 3.75 degrees for each 
“bin”. The accuracy of TID for a particular target is dominated 
by whether the measured angle is close to the centre of a TID 
“bin” or if it is close to the boundary between two “bins”. If the 
target phase angle is close to the boundary, its TID is more 
likely to icker between two values, depending on orientation, 
proximity to the coil etc. 
Q & A with Laurence
3 kHz - modied TID 
“bin” separation 
18.75 kHz - modied TID 
“bin” separation 
7.5 kHz- standard TID 
“bin” separation 
48
2 2 2
4
4
4
48
46
48
46
46
48










