D-6 Specifications
Specication for Actiwatch-L
ENVIRONMENTAL
Operating Storage
Temperature
0° C to 40° C
(32° F to 104° F)
-10° C to 50° C
(14° F to 122° F)
Relative Humidity
0 to 95% RH 0 to 95% RH
Atmospheric Pressure 500 to 1060kPa
(5600 feet to sea level)
PHYSICAL
Dimensions:
29 L x 37 W x 11 H (mm)
Weight:
16 grams w/o band
22 grams with standard band
Case Material: Glass-filled Nylon
Frame and Battery Cover: Titanium
Standard Wrist Band: Nylon with buckle
ELECTRICAL
Battery Type: CR 2025
Battery Life: 3 Months Typical
Protection Against Electric
Shock:
Internally powered
Degree of Protection Against
Electric Shock:
Type BF Applied Part
Modes of Operation: Continuous
Electromagnetic
Compatibility:
The device meets the
requirements of EN 60601-1-2, second
edition (2001).
Fuses: There are no user-replaceable fuses.
Degree of Protection Against
Ingress of Water:
Waterproof / Watertight to a depth of 1
meter for 30 minutes
IEC 60529 IPX7
DISPOSAL
When necessary, dispose of the device and accessories in accordance with local regulations.