User's Manual

E-2 EMC Information
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
±6V contact
±8V air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
Transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input/
output lines
±2kV for supply
mains
±1kV for input/
output lines
Mains power quality should be that of a
typical home or hospital environment.
It is noted certification to IEC 60601-
1-2isobtainedrecognizingthat
interruption of the download of data to
the reader due to EMI is not considered
to cause harm. When a download is
stopped, the user is prompted that
and error has occurred and the user
would then reinitiate the download.
Interference that would interrupt a
download is anticipated to be of short
duration such that the subsequent
download would be successful.
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common
mode
±1kV differential
mode
±2kV common
mode
Mains power quality should be that of a
typical home or hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30%
dip in U
T
) for 25
cycles
<5% U
T
(>95%
dip
in U
T
) for 5 sec
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30%
dip in U
T
) for 25
cycles
<5% U
T
(>95%
dip in U
T
) for 5
sec
Mains power quality should be that of a
typical home or hospital environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.