User Manual

10
8.2 Electrical Specifications
Accuracy: ± (a% reading + b digits), at 23°C ± 5°C, RH < 80%.
Range Accuracy Resolution
Test
Frequency
Maximum
Indication
200pF ±(0.5%+ 7) 0.1pF 800Hz 199.9pF
2nF
±(0.5%+ 5)
1pF 800Hz 1.999nF
20nF 10pF 800Hz 19.99nF
200nF 100µF 800Hz 199.9nF
2µF 1000µF 800Hz 1.999µF
20µF 0.01µF 80hz 19.99µF
200µF 0.1µF 8Hz 199.9µF
2000µF ±(2.0%+5) 1µF 8Hz 1999µF
20000µF ±(3.0%+10) 10µF 8Hz 1999(x 10)µF
pF = Pico Farad (10
-12
F)
nF = Nano Farad (10
-9
F)
µF = Micro Farad (10
-6
F)
Zero Error: 20pF
Test Voltage: 2.8V RMS maximum
Overload Protection: 0.1A/250V fuse.