Instruction manual
CHAPTER 2 : OPERATION GUIDE
126
Other
Functions
■ User Calibration dialog box (When the CS-2000/2000A is connected)
Lens Calibration
When registering a correction factor while using the optionally-available close-up lens CS-A35,
select Close up from the combo box.
Check
Click this button to display the correction factor for the lens specified in the combo box (i.e. the cor-
rection factor that is currently registered in the instrument).
Write
Click this button to display an Open dialog box. The file specified in this dialog box (file extension:
.lcf) is loaded, and then downloaded to the instrument.
The lens correction factor should be a text file of 401 pieces of data.
ND Calibration
In the combo box, select either ND1/10 or ND1/100.
When registering data for the optionally-available ND filter CS-A33 (1/10), select ND1/10. When
registering data for the optionally-available ND filter CS-A34 (1/100), select ND1/100.
Check
Click this button to display the correction factor for the ND filter specified in the combo box (i.e. the
correction factor that is currently registered in the instrument).
Write
Click this button to display an Open dialog box. The file specified in this dialog box (file extension:
.ncf) is loaded, and then downloaded to the instrument.
The correction factor for the ND filter should be a text file of 401 pieces of data.