Instruction manual
CHAPTER 2 : OPERATION GUIDE
127
Other
Functions
■ User Calibration dialog box (When the CS-200 is connected)
Lens/ND correction
Lens/ND
These check boxes can be edited only after you specify a file by using the Load button. Check an
appropriate box to select whether to use the value as a correction factor.
Load
Click this button to load the correction factor file.
A correction factor for a lens should be a text file of data of 40 sensors, and a correction factor for
the ND filter should be a text file of 401 pieces of data.
The correction factor files for the optional close-up lenses No. 107 and No. 122 are stored in the CS-
S10w with names “107.lcf ” and “122.lcf ”. (Refer to page 103.)
Clear
This button is effective when any file is referenced. Clicking this button clears the specified file.
Select field
Select one from User 1, User 2 and User 3.
Comment
Up to 18 characters can be used.
Lens/ND
The corresponding check box is checked to indicate whether the correction factor for a lens or an
ND filter is used.
Write
When this button is clicked, the correction factor is loaded to the instrument.
Clear
This button becomes effective after a correction factor is registered for the lens type specified in the
combo box. When this button is clicked, the correction factor loaded in the instrument is cleared.