CHAPTER 2 : OPERATION GUIDE
30
Starting the
CS-S10w
Software
When selecting Manual, set the seconds value in the box on the right between 0.0050000 and
120.0000 in increments of 0.000001 seconds.
You will also be able to select the CS-2000A's built-in ND filter setting from ON or OFF when Man-
ual is selected, or from ON, OFF or AUTO when the other option is selected.
Calibration CH
The combo box shows calibration channels in the form of “CH number: ID”. Select a calibration
channel between CH00 and CH10.
The calibration mode for the selected calibration channel is shown in the check boxes below. The
calibration mode is displayed as either Correct Wave Length or Correct Level, or both.
Instrument Settings dialog box (When the CS-200 is connected)
Measurement Type
Radiometric is displayed when the document type is set to the light-source color measurement.
[Selection disabled.]
When the document type is set to the object color measurement, select either Reflectance or Trans-
mittance.
Lens
Select the setting from Standard, Close up 1 (No. 107) and Close up 2 (No. 122). Select Close up 1
(No. 107) when close-up lens No. 107 (optional accessory) is attached, and select Close up 2 (No.
122) when close-up lens No. 122 (optional accessory) is attached.
When another lens such as an ND filter is attached and its correction factor is registered, User 1,
User 2 and User 3 are added to the options. To register correction factors, refer to page 124.
Measurement Angle
Either 1°, 0.2° or 0.1° is displayed depending on the status of the instrument. [Selection disabled]
Measurement Speed
Select the setting from Auto, LTD. AUTO, S-Fast, Fast, Slow, S-Slow and Manual. When selecting
Manual, enter the number of seconds between 1 and 60 in increments of 1 second in the box on the
right.
Measurement Mode
Select either Internal or No-Sync. When selecting Internal, set the Hz value in the box on the right
between 40.00 and 200.00 in increments of 0.01 Hz.
Calibration CH
The combo box shows calibration channels in the form of “CH number : ID name”. Select a calibra-
tion channel between CH00 and CH20.
The calibration mode for the selected calibration channel is shown in the text box on the right. The
calibration mode is Default or Single.