Instruction manual

CHAPTER 2 : OPERATION GUIDE
35
Preparing for
Measurement
Observer And Illuminant dialog box
Observer
Select either 2 degree or 10 degree.
Measurement illuminant
When the light-source color mode is selected
You can specify the desired white point which is used to calculate colorimetric data.
The available white point range is as follows:
X0/Y0/Z0: 0.01 to 99999000000.00
When the object color mode is selected
Select one from the following:
A, B, C, D50, D55, D65, D75, D95, E, F2, F6, F7, F8, F10, F11, F12, User 1, User 2 or User 3.
When selecting one of User 1 to 3, specify the user illuminant data file to be used. When the Set
button is clicked, the Input illuminant data dialog is displayed.
The setting in this dialog box will be reflected in all data included in the document file.
When the observer or illuminant is changed, the CS-S10w software recalculates all data. When you
attempt to change the observer or illuminant, the following message appears.
Input illuminant data dialog box
Tag
A tag of 60 characters or less can be specified for the illuminant data.